Sources of artifacts in rotational scanography

R. Moore, D. Korbuly, F. Kotula, W. Castaneda-Zuniga, K. Amplatz

Resultado de la investigación: Articlerevisión exhaustiva

Resumen

The images produced in rotational scanography, currently under investigation at several institutions, may contain several kinds of artifacts. The sources of such artifacts are: vertical and horizontal misalignment of the focal spot with the axis of rotation; misalignment of the primary and secondary slits; a density gradient due to slit cut-off; vibration of the x-ray tube during rotation; a density gradient due to variation in motor speed; radiographic striations due to ripple of the tube output. Striations are present even if a 12-pulse generator is used, but they disappear if a constant potential generator is employed.

Idioma originalEnglish (US)
Páginas (desde-hasta)333-340
Número de páginas8
PublicaciónFortschritte auf den Gebiete der Rontgenstrahlen und der Nuklearmedizin
Volumen137
N.º3
EstadoPublished - dic 1 1982