T1 - Reduction of the platelet review rate using the two-dimensional platelet method
AU - Chapman, Deanne H.
AU - Hardin, Jo Ann
AU - Miers, Martha
AU - Moyle, Susan
AU - Kinney, Marsha C.
PY - 2001
Y1 - 2001
N2 - Accurate platelet enumeration is critical for optimal treatment of patients with platelet and bleeding disorders, leukemias, and other neoplasias. The majority of automated hematology analyzers count platelets by size differentiation alone, which may result in falsely elevated platelet counts for samples containing interfering particles such as RBC fragments, microcytes, and cell debris. Most analyzers flag questionable platelet counts, necessitating review of results with confirmation by an alternative method, thus increasing the cost of performing platelet counts and delaying results. We studied the effect of a new platelet analysis method, based on measurement of size and refractive index, on the laboratory review rate for platelet counting. We demonstrated that this method yields higher accuracy for platelet counts in samples with interferences, especially for platelet counts less than 50 × 103/μL (<50 × 109/L). As a result of the 2-dimensional analysis, the review rate for platelet counts was reduced by 65% in our institution, resulting in substantial savings.
AB - Accurate platelet enumeration is critical for optimal treatment of patients with platelet and bleeding disorders, leukemias, and other neoplasias. The majority of automated hematology analyzers count platelets by size differentiation alone, which may result in falsely elevated platelet counts for samples containing interfering particles such as RBC fragments, microcytes, and cell debris. Most analyzers flag questionable platelet counts, necessitating review of results with confirmation by an alternative method, thus increasing the cost of performing platelet counts and delaying results. We studied the effect of a new platelet analysis method, based on measurement of size and refractive index, on the laboratory review rate for platelet counting. We demonstrated that this method yields higher accuracy for platelet counts in samples with interferences, especially for platelet counts less than 50 × 103/μL (<50 × 109/L). As a result of the 2-dimensional analysis, the review rate for platelet counts was reduced by 65% in our institution, resulting in substantial savings.
KW - Flow cytometry
KW - Laboratory review rate
KW - Light scatter
KW - Optical platelet analysis
KW - Platelet count
KW - RBC interference
UR - http://www.scopus.com/inward/record.url?scp=0035001115&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0035001115&partnerID=8YFLogxK
U2 - 10.1309/C563-49VA-JYDU-LEJG
DO - 10.1309/C563-49VA-JYDU-LEJG
M3 - Article
C2 - 11392887
AN - SCOPUS:0035001115
SN - 0002-9173
VL - 115
SP - 894
EP - 898
JO - American journal of clinical pathology
JF - American journal of clinical pathology
IS - 6
ER -