Layer thickness in panoramic radiography as defined by different noise-equivalent passbands

Masaru Shiojima, Åke Bäckström, Ulf Welander, William D. McDavid, Gunilla Tronje, Munetaka Naitoh

Resultado de la investigación: Articlerevisión exhaustiva

5 Citas (Scopus)

Resumen

The layer thickness in rotational panoramic radiography is presented with the use of the concept of the noise-equivalent passband. Conventionally, the layer thickness has been calculated only from one-dimensional data in the horizontal or rotational dimension of panoramic radiography. In the present study, results from calculations using both one- and two-dimensional data are presented. When the vertical dimension is included in the calculations, the layer is wider than when calculated from data in the horizontal dimension only. It is pointed out that the wider layer that follows from the introduction of the vertical dimension may be the most reliable measure of layer thickness in panoramic radiography.

Idioma originalEnglish (US)
Páginas (desde-hasta)244-250
Número de páginas7
PublicaciónOral Surgery, Oral Medicine, Oral Pathology
Volumen76
N.º2
DOI
EstadoPublished - ago. 1993