High-resolution CT of the lungs: An optimal approach

J. R. Mayo, W. R. Webb, R. Gould, M. G. Stein, I. Bass, G. Gamsu, H. I. Goldberg

Resultado de la investigación: Articlerevisión exhaustiva

203 Citas (Scopus)

Resumen

The influences of kilovolt peak, milliamperage, reconstruction algorithm, targeting, and image magnification on thin-section (1.5-mm) computed tomography (CT) of the lungs were studied in phantoms and patients. Retrospective targeted reconstruction (25-cm field of view) improved spatial resolution, while magnification did not. The bone reconstruction algorithm improved spatial resolution, compared with the standard algorithm, and in patients, bone algorithm images were considered superior to standard reconstructions. Although using the bone algorithm increases the visible image noise, increasing the kilovolt peak and the milliamperage can reduce this noise. However, in the patients studied, this reduction in noise was not usually judged as significant, except in the posterior, paravertebral part of the lung. An optimal technique for CT of the lung parenchyma should include thin-collimation, targeted scans reconstructed with a high-spatial frequency algorithm and, in some patients, increased kilovolt peak or milliamperage.

Idioma originalEnglish (US)
Páginas (desde-hasta)507-510
Número de páginas4
PublicaciónRadiology
Volumen163