Ir directamente a la navegación principal Ir directamente a la búsqueda Ir directamente al contenido principal

Field-Ready Suction Solutions for Emergencies: The Battlefield Ready Innovative Suction Kit (BRISK)

Producción científica: Article › revisión exhaustiva

Resumen

Introduction: Portable suction devices are crucial for emergency airway management. Commercially-available units are unsuitable for field use due to size and power needs. A light-weight and multi-orientation operable portable suction device, Battlefield Ready Innovative Suction Kit (BRISK) was developed. The design was informed by feedback from combat medics, paramedics, and EMTs. Methods: End-user engagement and feedback defined BRISK’s design. The fabricated prototype used a vacuum pump and hydrophobic syringe filters. Performance tests measured vacuum pressure, air and liquid (water and ISO vomit simulant) flow rates, volume of water suctioned in different orientations (upright, tilted, or inverted), and contamination prevention between BRISK, SSCOR Quickdraw, and Laerdal LCSU4. Results: The BRISK device—weighing 0.97 kg—demonstrated a maximum vacuum pressure of 570 ± 6 mmHg and an air flowrate of 5.20 L/min. Liquid flow rates (L/min) for BRISK, LCSU4, and SSCOR with water were 4.92 ± 0.2, 6.97 ± 0.1, and 5.37 ± 0.1, respectively. With ISO vomit simulant, the rates were 3.23 ± 0.2, 3.06 ± 0.4, and 2.23 ± 0.1. BRISK showed consistent performance across orientations (p = 0.081), while LCSU4 and SSCOR varied significantly (p < 0.0001). The BRISK’s cross-contamination between filters and the pump was 0.01%, which is far less than the rated contamination level of 0.1% as described by ISO 10079-1. Conclusion: BRISK is 30% lighter and achieves competitive vacuum pressures and liquid flow rates than comparable units currently in the market. It ensures effective evacuation of liquids in all orientations: upright, tilted, or inverted. These results demonstrate the BRISK has the potential to provide superior clinical performance in pre-hospital and first response scenarios.

Idioma originalEnglish (US)
Número de artículo104211
Páginas (desde-hasta)1409-1422
Número de páginas14
PublicaciónAnnals of Biomedical Engineering
Volumen53
N.º6
DOI
EstadoPublished - jun 2025

ASJC Scopus subject areas

  • Biomedical Engineering

Huella

Profundice en los temas de investigación de 'Field-Ready Suction Solutions for Emergencies: The Battlefield Ready Innovative Suction Kit (BRISK)'. En conjunto forman una huella única.

Citar esto