Effect of electron contamination on in vivo dosimetry for lung block shielding during TBI

Ganesh Narayanasamy, Wilbert Cruz, Daniel L. Saenz, Sotirios Stathakis, Niko Papanikolaou, Neil Kirby

Resultado de la investigación: Articlerevisión exhaustiva

6 Citas (Scopus)

Resumen

Our institution performs in vivo verification measurement for each of our total body irradiation (TBI) patients with optically stimulated luminescent dosimeters (OSLD). The lung block verification measurements were commonly higher than expected. The aim of this work is to understand this discrepancy and improve the accuracy of these lung block verification measurements. Initially, the thickness of the lung block was increased to provide adequate lung sparing. Further tests revealed the increase was due to electron contamination dose emanating from the lung block. The thickness of the bolus material covering the OSLD behind the lung block was increased to offset the electron contamination. In addition, the distance from the lung block to the dosimeter was evaluated for its effect on the OSLD reading and found to be clinically insignificant over the range of variability in our clinic. The results show that the improved TBI treatment technique provides for better accuracy of measured dose in vivo and consistency of patient setup.

Idioma originalEnglish (US)
Páginas (desde-hasta)486-491
Número de páginas6
PublicaciónJournal of Applied Clinical Medical Physics
Volumen17
N.º3
DOI
EstadoPublished - 2016

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Huella

Profundice en los temas de investigación de 'Effect of electron contamination on in vivo dosimetry for lung block shielding during TBI'. En conjunto forman una huella única.

Citar esto