Dosimetric evaluation of PLATO and Oncentra treatment planning systems for High Dose Rate (HDR) brachytherapy gynecological treatments

Hardev Singh, Tania De La Fuente Herman, Barry Showalter, Spencer J. Thompson, Elizabeth J. Syzek, Terence Herman, Salahuddin Ahmad

Resultado de la investigación: Conference contribution

1 Cita (Scopus)

Resumen

This study compares the dosimetric differences in HDR brachytherapy treatment plans calculated with Nucletron's PLATO and Oncentra MasterPlan treatment planning systems (TPS). Ten patients (1 T1b, 1 T2a, 6 T2b, 2 T4) having cervical carcinoma, median age of 43.5 years (range, 34-79 years) treated with tandem & ring applicator in our institution were selected retrospectively for this study. For both Plato and Oncentra TPS, the same orthogonal films anterior-posterior (AP) and lateral were used to manually draw the prescription and anatomical points using definitions from the Manchester system and recommendations from the ICRU report 38. Data input for PLATO was done using a digitizer and Epson Expression 10000XL scanner was used for Oncentra where the points were selected on the images in the screen. The prescription doses for these patients were 30 Gy to points right A (RA) and left A (LA) delivered in 5 fractions with Ir-192 HDR source. Two arrangements: one dwell position and two dwell positions on the tandem were used for dose calculation. The doses to the patient points right B (RB) and left B (LB), and to the organs at risk (OAR), bladder and rectum for each patient were calculated. The mean dose and the mean percentage difference in dose calculated by the two treatment planning systems were compared. Paired t-tests were used for statistical analysis. No significant differences in mean RB, LB, bladder and rectum doses were found with p-values > 0.14. The mean percent difference of doses in RB, LB, bladder and rectum are found to be less than 2.2%, 1.8%, 1.3% and 2.2%, respectively. Dose calculations based on the two different treatment planning systems were found to be consistent and the treatment plans can be made with either system in our department without any concern.

Idioma originalEnglish (US)
Título de la publicación alojadaMedical Physics - Twelfth Mexican Symposium on Medical Physics
Páginas70-72
Número de páginas3
DOI
EstadoPublished - 2012
Publicado de forma externa
Evento12th Mexican Symposium on Medical Physics - Oaxaca, Mexico
Duración: mar 16 2012mar 19 2012

Serie de la publicación

NombreAIP Conference Proceedings
Volumen1494
ISSN (versión impresa)0094-243X
ISSN (versión digital)1551-7616

Conference

Conference12th Mexican Symposium on Medical Physics
País/TerritorioMexico
CiudadOaxaca
Período3/16/123/19/12

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Huella

Profundice en los temas de investigación de 'Dosimetric evaluation of PLATO and Oncentra treatment planning systems for High Dose Rate (HDR) brachytherapy gynecological treatments'. En conjunto forman una huella única.

Citar esto