TY - JOUR
T1 - Agreement Between Swept-source Optical Coherence Tomography and Rotating Scheimpflug Camera in Measurement of Corneal Parameters in Normal and Keratoconic Eyes
AU - Ghassemi, Hamed
AU - Zarei-Ghanavati, Mehran
AU - Khastavan, Mina
AU - Atighehchian, Mehrnaz
AU - Khorasani, Abbas Azimi
AU - Latifi, Golshan
N1 - Publisher Copyright:
© 2023 Atighehchian et al.
PY - 2023/10
Y1 - 2023/10
N2 - Purpose: This study aimed to assess the agreement between topographic indices of healthy subjects and keratoconus (KCN) patients using a swept-source optical coherence tomography (SS-OCT CASIA2) versus a Scheimpflug camera (Pentacam). Methods: 40 eyes of 23 patients with KCN and 40 eyes of 20 healthy subjects were included and evaluated with the CASIA2, followed by the Pentacam. Two consecutive modalities were obtained for one eye of each patient. Corneal parameters, including anterior keratometry at steep (Ks) and flat meridians (Kf), anterior astigmatism, anterior and posterior corneal elevation values, thinnest corneal thickness, and apex corneal thickness, were evaluated. Results: CASIA2 and Pentacam showed perfect agreement (95% limits of agreement (LoA): -0.22 to 0.68, 95% LoA: -1.5 to 1.44 D) and good correlation (Intraclass correlation (ICC):0.986, ICC:0.987; P <0.01) for anterior (Ks) in normal and ectatic corneas, respectively. The cylinder amount had moderate agreement and correlation (95% LoA: -0.55 to 0.47D, ICC: 0.797, P <0.01) in normal, and moderate to strong agreement and correlation (95% LoA: -1.57 to 0.87D, ICC=0.911, P <0.01) in Keratoconic eyes. There was a fair agreement for anterior and posterior corneal elevation values in normal subjects (95% LoA: -3.09 to 4.59, 95% LoA: -6.91 to 7.31D). The thinnest corneal thickness amount had an excellent agreement in normal and KCN patients (ICC: 0.983, 0.953; respectively). Conclusions: Although the devices had different mean indices values, they had a good agreement based on the Bland-Altman plots. Since Pentacam is accepted as the standard tool for diagnosing ectatic cornea, pentacam CASIA2 is also helpful for early diagnosis of KCN.
AB - Purpose: This study aimed to assess the agreement between topographic indices of healthy subjects and keratoconus (KCN) patients using a swept-source optical coherence tomography (SS-OCT CASIA2) versus a Scheimpflug camera (Pentacam). Methods: 40 eyes of 23 patients with KCN and 40 eyes of 20 healthy subjects were included and evaluated with the CASIA2, followed by the Pentacam. Two consecutive modalities were obtained for one eye of each patient. Corneal parameters, including anterior keratometry at steep (Ks) and flat meridians (Kf), anterior astigmatism, anterior and posterior corneal elevation values, thinnest corneal thickness, and apex corneal thickness, were evaluated. Results: CASIA2 and Pentacam showed perfect agreement (95% limits of agreement (LoA): -0.22 to 0.68, 95% LoA: -1.5 to 1.44 D) and good correlation (Intraclass correlation (ICC):0.986, ICC:0.987; P <0.01) for anterior (Ks) in normal and ectatic corneas, respectively. The cylinder amount had moderate agreement and correlation (95% LoA: -0.55 to 0.47D, ICC: 0.797, P <0.01) in normal, and moderate to strong agreement and correlation (95% LoA: -1.57 to 0.87D, ICC=0.911, P <0.01) in Keratoconic eyes. There was a fair agreement for anterior and posterior corneal elevation values in normal subjects (95% LoA: -3.09 to 4.59, 95% LoA: -6.91 to 7.31D). The thinnest corneal thickness amount had an excellent agreement in normal and KCN patients (ICC: 0.983, 0.953; respectively). Conclusions: Although the devices had different mean indices values, they had a good agreement based on the Bland-Altman plots. Since Pentacam is accepted as the standard tool for diagnosing ectatic cornea, pentacam CASIA2 is also helpful for early diagnosis of KCN.
KW - CASIA2
KW - Keratoconus
KW - Optical Coherence Tomography
KW - Pentacam
UR - http://www.scopus.com/inward/record.url?scp=85183513046&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85183513046&partnerID=8YFLogxK
U2 - 10.18502/jovr.v18i4.14547
DO - 10.18502/jovr.v18i4.14547
M3 - Article
C2 - 38250237
AN - SCOPUS:85183513046
SN - 2008-2010
VL - 18
SP - 386
EP - 395
JO - Journal of Ophthalmic and Vision Research
JF - Journal of Ophthalmic and Vision Research
IS - 4
ER -