A modified technique to minimize filter tilting during deployment of the Günther Tulip filter: In vitro study

Jorge E. Lopera, Junko U. Araki, David Kirsch, Zhong Qian, Augusto Brazzini, Arturo Gonzalez, Wilfrido Castaneda-Zuniga

Producción científica: Articlerevisión exhaustiva

17 Citas (Scopus)

Resumen

With an in vitro model of the inferior vena cava, it was demonstrated that slight tension on the introducer during quick release of the jugular Günther Tulip filter led to less tilting of the filter compared with the technique recommended by the manufacturer. This technique may be useful to minimize significant filter tilting that may be associated with decreased filtration efficiency, and difficulty or impossibility of future filter retrieval.