Time-intensity-based volumetric flow measurements: An in vitro study

P. C. Li, Chih-ko Yeh, S. W. Wang

Research output: Contribution to journalArticle

22 Citations (Scopus)

Abstract

Ultrasonic contrast agents have been used to assist blood flow measurements. Several contrast-specific flow measurement techniques have been proposed during the last few years. Among them, a method based on relative enhancement of the backscattered signal as a function of time is of particular interest. This method is also known as the time-intensity method. The method is based on the indicator-dilution theory, and the time-intensity curve is used to derive blood flow-related parameters such as the flow rate and the blood mixing volume. Previous in vitro studies done by other research groups were mainly based on a perfusion model or an artery model. Results showed that several parameters derived from the time-intensity curve had a good correlation with the flow rate under certain conditions. However, the studies did not focus on factors such as mixing volume, mixing chamber configuration and different types of mixing chamber. In this paper, dependence of the time-intensity curve is further studied. Specifically, two types of blood-mixing chambers were constructed. One was a spherical compartment phantom with two different sizes (260 and 580 mL) and different inflow/outflow configurations. The other was a perfusion phantom consisting of dialysis cartridges with the volume ranging from 114 to 351 mL. The time intensities were also measured at both the input and the output of the mixing chamber. A commercial agent (Levovist®) and a self-made, albumin-based agent were used and the wash-out time constant and the mean transit time were derived for flow rates ranging from 500 to 1300 mL/min. For the perfusion phantom, results showed that the parameters had a good correlation with both the flow rate and the mixing volume. Results from the compartment phantom, on the other hand, indicated that the inflow/outflow configuration and the mixing size significantly affected the derived time constants. Potential applications of new volumetric flow estimation techniques based on both input and output intensities were also discussed.

Original languageEnglish (US)
Pages (from-to)349-358
Number of pages10
JournalUltrasound in Medicine and Biology
Volume28
Issue number3
DOIs
StatePublished - 2002
Externally publishedYes

Fingerprint

flow measurement
flow velocity
chambers
compartments
blood flow
Perfusion
time constant
blood
curves
configurations
cartridges
dialysis
In Vitro Techniques
output
transit time
arteries
albumins
Blood Volume
dilution
Ultrasonics

Keywords

  • Time-intensity curve
  • Ultrasonic contrast agent
  • Volumetric flow measurement

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Time-intensity-based volumetric flow measurements : An in vitro study. / Li, P. C.; Yeh, Chih-ko; Wang, S. W.

In: Ultrasound in Medicine and Biology, Vol. 28, No. 3, 2002, p. 349-358.

Research output: Contribution to journalArticle

@article{20d0b9c9cb554d4b8d4598cee7e3545a,
title = "Time-intensity-based volumetric flow measurements: An in vitro study",
abstract = "Ultrasonic contrast agents have been used to assist blood flow measurements. Several contrast-specific flow measurement techniques have been proposed during the last few years. Among them, a method based on relative enhancement of the backscattered signal as a function of time is of particular interest. This method is also known as the time-intensity method. The method is based on the indicator-dilution theory, and the time-intensity curve is used to derive blood flow-related parameters such as the flow rate and the blood mixing volume. Previous in vitro studies done by other research groups were mainly based on a perfusion model or an artery model. Results showed that several parameters derived from the time-intensity curve had a good correlation with the flow rate under certain conditions. However, the studies did not focus on factors such as mixing volume, mixing chamber configuration and different types of mixing chamber. In this paper, dependence of the time-intensity curve is further studied. Specifically, two types of blood-mixing chambers were constructed. One was a spherical compartment phantom with two different sizes (260 and 580 mL) and different inflow/outflow configurations. The other was a perfusion phantom consisting of dialysis cartridges with the volume ranging from 114 to 351 mL. The time intensities were also measured at both the input and the output of the mixing chamber. A commercial agent (Levovist{\circledR}) and a self-made, albumin-based agent were used and the wash-out time constant and the mean transit time were derived for flow rates ranging from 500 to 1300 mL/min. For the perfusion phantom, results showed that the parameters had a good correlation with both the flow rate and the mixing volume. Results from the compartment phantom, on the other hand, indicated that the inflow/outflow configuration and the mixing size significantly affected the derived time constants. Potential applications of new volumetric flow estimation techniques based on both input and output intensities were also discussed.",
keywords = "Time-intensity curve, Ultrasonic contrast agent, Volumetric flow measurement",
author = "Li, {P. C.} and Chih-ko Yeh and Wang, {S. W.}",
year = "2002",
doi = "10.1016/S0301-5629(01)00516-6",
language = "English (US)",
volume = "28",
pages = "349--358",
journal = "Ultrasound in Medicine and Biology",
issn = "0301-5629",
publisher = "Elsevier USA",
number = "3",

}

TY - JOUR

T1 - Time-intensity-based volumetric flow measurements

T2 - An in vitro study

AU - Li, P. C.

AU - Yeh, Chih-ko

AU - Wang, S. W.

PY - 2002

Y1 - 2002

N2 - Ultrasonic contrast agents have been used to assist blood flow measurements. Several contrast-specific flow measurement techniques have been proposed during the last few years. Among them, a method based on relative enhancement of the backscattered signal as a function of time is of particular interest. This method is also known as the time-intensity method. The method is based on the indicator-dilution theory, and the time-intensity curve is used to derive blood flow-related parameters such as the flow rate and the blood mixing volume. Previous in vitro studies done by other research groups were mainly based on a perfusion model or an artery model. Results showed that several parameters derived from the time-intensity curve had a good correlation with the flow rate under certain conditions. However, the studies did not focus on factors such as mixing volume, mixing chamber configuration and different types of mixing chamber. In this paper, dependence of the time-intensity curve is further studied. Specifically, two types of blood-mixing chambers were constructed. One was a spherical compartment phantom with two different sizes (260 and 580 mL) and different inflow/outflow configurations. The other was a perfusion phantom consisting of dialysis cartridges with the volume ranging from 114 to 351 mL. The time intensities were also measured at both the input and the output of the mixing chamber. A commercial agent (Levovist®) and a self-made, albumin-based agent were used and the wash-out time constant and the mean transit time were derived for flow rates ranging from 500 to 1300 mL/min. For the perfusion phantom, results showed that the parameters had a good correlation with both the flow rate and the mixing volume. Results from the compartment phantom, on the other hand, indicated that the inflow/outflow configuration and the mixing size significantly affected the derived time constants. Potential applications of new volumetric flow estimation techniques based on both input and output intensities were also discussed.

AB - Ultrasonic contrast agents have been used to assist blood flow measurements. Several contrast-specific flow measurement techniques have been proposed during the last few years. Among them, a method based on relative enhancement of the backscattered signal as a function of time is of particular interest. This method is also known as the time-intensity method. The method is based on the indicator-dilution theory, and the time-intensity curve is used to derive blood flow-related parameters such as the flow rate and the blood mixing volume. Previous in vitro studies done by other research groups were mainly based on a perfusion model or an artery model. Results showed that several parameters derived from the time-intensity curve had a good correlation with the flow rate under certain conditions. However, the studies did not focus on factors such as mixing volume, mixing chamber configuration and different types of mixing chamber. In this paper, dependence of the time-intensity curve is further studied. Specifically, two types of blood-mixing chambers were constructed. One was a spherical compartment phantom with two different sizes (260 and 580 mL) and different inflow/outflow configurations. The other was a perfusion phantom consisting of dialysis cartridges with the volume ranging from 114 to 351 mL. The time intensities were also measured at both the input and the output of the mixing chamber. A commercial agent (Levovist®) and a self-made, albumin-based agent were used and the wash-out time constant and the mean transit time were derived for flow rates ranging from 500 to 1300 mL/min. For the perfusion phantom, results showed that the parameters had a good correlation with both the flow rate and the mixing volume. Results from the compartment phantom, on the other hand, indicated that the inflow/outflow configuration and the mixing size significantly affected the derived time constants. Potential applications of new volumetric flow estimation techniques based on both input and output intensities were also discussed.

KW - Time-intensity curve

KW - Ultrasonic contrast agent

KW - Volumetric flow measurement

UR - http://www.scopus.com/inward/record.url?scp=0036225723&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0036225723&partnerID=8YFLogxK

U2 - 10.1016/S0301-5629(01)00516-6

DO - 10.1016/S0301-5629(01)00516-6

M3 - Article

C2 - 11978415

AN - SCOPUS:0036225723

VL - 28

SP - 349

EP - 358

JO - Ultrasound in Medicine and Biology

JF - Ultrasound in Medicine and Biology

SN - 0301-5629

IS - 3

ER -