The level of detail required in a deformable phantom to accurately perform quality assurance of deformable image registration

Daniel L. Saenz, Hojin Kim, Josephine Chen, Sotirios Stathakis, Neil Kirby

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

The primary purpose of the study was to determine how detailed deformable image registration (DIR) phantoms need to adequately simulate human anatomy and accurately assess the quality of DIR algorithms. In particular, how many distinct tissues are required in a phantom to simulate complex human anatomy? Pelvis and head-and-neck patient CT images were used for this study as virtual phantoms. Two data sets from each site were analyzed. The virtual phantoms were warped to create two pairs consisting of undeformed and deformed images. Otsu's method was employed to create additional segmented image pairs of n distinct soft tissue CT number ranges (fat, muscle, etc). A realistic noise image was added to each image. Deformations were applied in MIM Software (MIM) and Velocity deformable multi-pass (DMP) and compared with the known warping. Images with more simulated tissue levels exhibit more contrast, enabling more accurate results. Deformation error (magnitude of the vector difference between known and predicted deformation) was used as a metric to evaluate how many CT number gray levels are needed for a phantom to serve as a realistic patient proxy. Stabilization of the mean deformation error was reached by three soft tissue levels for Velocity DMP and MIM, though MIM exhibited a persisting difference in accuracy between the discrete images and the unprocessed image pair. A minimum detail of three levels allows a realistic patient proxy for use with Velocity and MIM deformation algorithms.

Original languageEnglish (US)
Pages (from-to)6269-6280
Number of pages12
JournalPhysics in Medicine and Biology
Volume61
Issue number17
DOIs
StatePublished - Aug 5 2016

Fingerprint

Proxy
Anatomy
Pelvis
Noise
Neck
Software
Fats
Head
Muscles
Datasets

Keywords

  • deformable image registration
  • deformable phantoms
  • quality assurance

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Cite this

The level of detail required in a deformable phantom to accurately perform quality assurance of deformable image registration. / Saenz, Daniel L.; Kim, Hojin; Chen, Josephine; Stathakis, Sotirios; Kirby, Neil.

In: Physics in Medicine and Biology, Vol. 61, No. 17, 05.08.2016, p. 6269-6280.

Research output: Contribution to journalArticle

Saenz, Daniel L. ; Kim, Hojin ; Chen, Josephine ; Stathakis, Sotirios ; Kirby, Neil. / The level of detail required in a deformable phantom to accurately perform quality assurance of deformable image registration. In: Physics in Medicine and Biology. 2016 ; Vol. 61, No. 17. pp. 6269-6280.
@article{b14171487d364c678688f872bef5bb85,
title = "The level of detail required in a deformable phantom to accurately perform quality assurance of deformable image registration",
abstract = "The primary purpose of the study was to determine how detailed deformable image registration (DIR) phantoms need to adequately simulate human anatomy and accurately assess the quality of DIR algorithms. In particular, how many distinct tissues are required in a phantom to simulate complex human anatomy? Pelvis and head-and-neck patient CT images were used for this study as virtual phantoms. Two data sets from each site were analyzed. The virtual phantoms were warped to create two pairs consisting of undeformed and deformed images. Otsu's method was employed to create additional segmented image pairs of n distinct soft tissue CT number ranges (fat, muscle, etc). A realistic noise image was added to each image. Deformations were applied in MIM Software (MIM) and Velocity deformable multi-pass (DMP) and compared with the known warping. Images with more simulated tissue levels exhibit more contrast, enabling more accurate results. Deformation error (magnitude of the vector difference between known and predicted deformation) was used as a metric to evaluate how many CT number gray levels are needed for a phantom to serve as a realistic patient proxy. Stabilization of the mean deformation error was reached by three soft tissue levels for Velocity DMP and MIM, though MIM exhibited a persisting difference in accuracy between the discrete images and the unprocessed image pair. A minimum detail of three levels allows a realistic patient proxy for use with Velocity and MIM deformation algorithms.",
keywords = "deformable image registration, deformable phantoms, quality assurance",
author = "Saenz, {Daniel L.} and Hojin Kim and Josephine Chen and Sotirios Stathakis and Neil Kirby",
year = "2016",
month = "8",
day = "5",
doi = "10.1088/0031-9155/61/17/6269",
language = "English (US)",
volume = "61",
pages = "6269--6280",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "17",

}

TY - JOUR

T1 - The level of detail required in a deformable phantom to accurately perform quality assurance of deformable image registration

AU - Saenz, Daniel L.

AU - Kim, Hojin

AU - Chen, Josephine

AU - Stathakis, Sotirios

AU - Kirby, Neil

PY - 2016/8/5

Y1 - 2016/8/5

N2 - The primary purpose of the study was to determine how detailed deformable image registration (DIR) phantoms need to adequately simulate human anatomy and accurately assess the quality of DIR algorithms. In particular, how many distinct tissues are required in a phantom to simulate complex human anatomy? Pelvis and head-and-neck patient CT images were used for this study as virtual phantoms. Two data sets from each site were analyzed. The virtual phantoms were warped to create two pairs consisting of undeformed and deformed images. Otsu's method was employed to create additional segmented image pairs of n distinct soft tissue CT number ranges (fat, muscle, etc). A realistic noise image was added to each image. Deformations were applied in MIM Software (MIM) and Velocity deformable multi-pass (DMP) and compared with the known warping. Images with more simulated tissue levels exhibit more contrast, enabling more accurate results. Deformation error (magnitude of the vector difference between known and predicted deformation) was used as a metric to evaluate how many CT number gray levels are needed for a phantom to serve as a realistic patient proxy. Stabilization of the mean deformation error was reached by three soft tissue levels for Velocity DMP and MIM, though MIM exhibited a persisting difference in accuracy between the discrete images and the unprocessed image pair. A minimum detail of three levels allows a realistic patient proxy for use with Velocity and MIM deformation algorithms.

AB - The primary purpose of the study was to determine how detailed deformable image registration (DIR) phantoms need to adequately simulate human anatomy and accurately assess the quality of DIR algorithms. In particular, how many distinct tissues are required in a phantom to simulate complex human anatomy? Pelvis and head-and-neck patient CT images were used for this study as virtual phantoms. Two data sets from each site were analyzed. The virtual phantoms were warped to create two pairs consisting of undeformed and deformed images. Otsu's method was employed to create additional segmented image pairs of n distinct soft tissue CT number ranges (fat, muscle, etc). A realistic noise image was added to each image. Deformations were applied in MIM Software (MIM) and Velocity deformable multi-pass (DMP) and compared with the known warping. Images with more simulated tissue levels exhibit more contrast, enabling more accurate results. Deformation error (magnitude of the vector difference between known and predicted deformation) was used as a metric to evaluate how many CT number gray levels are needed for a phantom to serve as a realistic patient proxy. Stabilization of the mean deformation error was reached by three soft tissue levels for Velocity DMP and MIM, though MIM exhibited a persisting difference in accuracy between the discrete images and the unprocessed image pair. A minimum detail of three levels allows a realistic patient proxy for use with Velocity and MIM deformation algorithms.

KW - deformable image registration

KW - deformable phantoms

KW - quality assurance

UR - http://www.scopus.com/inward/record.url?scp=84984697582&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84984697582&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/61/17/6269

DO - 10.1088/0031-9155/61/17/6269

M3 - Article

C2 - 27494827

AN - SCOPUS:84984697582

VL - 61

SP - 6269

EP - 6280

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 17

ER -