SU‐GG‐T‐244: Commissioning of Flattening Filter Free Varian 600C Linear Accelerator

Sotirios Stathakis, A. Gutiérrez, P. Rassiah‐szegedi, P. Mavroidis, Nikos Papanikolaou

Research output: Contribution to journalArticle

Abstract

Purpose: The purpose of this study is twofold: a)to determine the dosimetric features of an unflattened 6‐MV photon beam of a Varian 600C linac; and b)to compare the beam characteristics of the unflattened beam against the flattened one. Method and Materials: Comparison of measurements in water was performed for a Varian 600C linear accelerator with and without flattening filter. The measurements consisted of dose profiles (inplane and cross plane), percent depth doses, and photon beam output. The field sizes measured varied from 2cm×2cm to 40cm×40cm. All measurements were taken at 100cm SSD with an A1SL ion chamber. Results: Our results showed a significant (about twofold) increase in the dose rate for all field sizes. Also, a steeper reduction in depth doses was observed when the flattening filter was removed. No decrease was observed in the out‐of‐field dose, but there was an increase of the dose in the build‐up region that can be attributed to the increase of contaminant electrons. Conclusion:. Our study revealed that, for recent radiotherapy techniques, removing the flattening filter could be beneficial since there is a an increased dose rate and higher output therefore less monitor are required for each fraction leading to reduced whole body dose equivalent. On the other hand, higher skin doses are observed per monitor unit delivered but the overall skin dose for a complete treatment is comparable to the one with flattening filter since less monitor unit are required per fraction.

Original languageEnglish (US)
Pages (from-to)2781
Number of pages1
JournalMedical Physics
Volume35
Issue number6
DOIs
StatePublished - 2008

Fingerprint

Particle Accelerators
Photons
Silver Sulfadiazine
Skin
Radiotherapy
Electrons
Ions
Water
Therapeutics

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

SU‐GG‐T‐244 : Commissioning of Flattening Filter Free Varian 600C Linear Accelerator. / Stathakis, Sotirios; Gutiérrez, A.; Rassiah‐szegedi, P.; Mavroidis, P.; Papanikolaou, Nikos.

In: Medical Physics, Vol. 35, No. 6, 2008, p. 2781.

Research output: Contribution to journalArticle

Stathakis, Sotirios ; Gutiérrez, A. ; Rassiah‐szegedi, P. ; Mavroidis, P. ; Papanikolaou, Nikos. / SU‐GG‐T‐244 : Commissioning of Flattening Filter Free Varian 600C Linear Accelerator. In: Medical Physics. 2008 ; Vol. 35, No. 6. pp. 2781.
@article{abbfd4511fa74bb3bb8c86d91b1683a1,
title = "SU‐GG‐T‐244: Commissioning of Flattening Filter Free Varian 600C Linear Accelerator",
abstract = "Purpose: The purpose of this study is twofold: a)to determine the dosimetric features of an unflattened 6‐MV photon beam of a Varian 600C linac; and b)to compare the beam characteristics of the unflattened beam against the flattened one. Method and Materials: Comparison of measurements in water was performed for a Varian 600C linear accelerator with and without flattening filter. The measurements consisted of dose profiles (inplane and cross plane), percent depth doses, and photon beam output. The field sizes measured varied from 2cm×2cm to 40cm×40cm. All measurements were taken at 100cm SSD with an A1SL ion chamber. Results: Our results showed a significant (about twofold) increase in the dose rate for all field sizes. Also, a steeper reduction in depth doses was observed when the flattening filter was removed. No decrease was observed in the out‐of‐field dose, but there was an increase of the dose in the build‐up region that can be attributed to the increase of contaminant electrons. Conclusion:. Our study revealed that, for recent radiotherapy techniques, removing the flattening filter could be beneficial since there is a an increased dose rate and higher output therefore less monitor are required for each fraction leading to reduced whole body dose equivalent. On the other hand, higher skin doses are observed per monitor unit delivered but the overall skin dose for a complete treatment is comparable to the one with flattening filter since less monitor unit are required per fraction.",
author = "Sotirios Stathakis and A. Guti{\'e}rrez and P. Rassiah‐szegedi and P. Mavroidis and Nikos Papanikolaou",
year = "2008",
doi = "10.1118/1.2961996",
language = "English (US)",
volume = "35",
pages = "2781",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐GG‐T‐244

T2 - Commissioning of Flattening Filter Free Varian 600C Linear Accelerator

AU - Stathakis, Sotirios

AU - Gutiérrez, A.

AU - Rassiah‐szegedi, P.

AU - Mavroidis, P.

AU - Papanikolaou, Nikos

PY - 2008

Y1 - 2008

N2 - Purpose: The purpose of this study is twofold: a)to determine the dosimetric features of an unflattened 6‐MV photon beam of a Varian 600C linac; and b)to compare the beam characteristics of the unflattened beam against the flattened one. Method and Materials: Comparison of measurements in water was performed for a Varian 600C linear accelerator with and without flattening filter. The measurements consisted of dose profiles (inplane and cross plane), percent depth doses, and photon beam output. The field sizes measured varied from 2cm×2cm to 40cm×40cm. All measurements were taken at 100cm SSD with an A1SL ion chamber. Results: Our results showed a significant (about twofold) increase in the dose rate for all field sizes. Also, a steeper reduction in depth doses was observed when the flattening filter was removed. No decrease was observed in the out‐of‐field dose, but there was an increase of the dose in the build‐up region that can be attributed to the increase of contaminant electrons. Conclusion:. Our study revealed that, for recent radiotherapy techniques, removing the flattening filter could be beneficial since there is a an increased dose rate and higher output therefore less monitor are required for each fraction leading to reduced whole body dose equivalent. On the other hand, higher skin doses are observed per monitor unit delivered but the overall skin dose for a complete treatment is comparable to the one with flattening filter since less monitor unit are required per fraction.

AB - Purpose: The purpose of this study is twofold: a)to determine the dosimetric features of an unflattened 6‐MV photon beam of a Varian 600C linac; and b)to compare the beam characteristics of the unflattened beam against the flattened one. Method and Materials: Comparison of measurements in water was performed for a Varian 600C linear accelerator with and without flattening filter. The measurements consisted of dose profiles (inplane and cross plane), percent depth doses, and photon beam output. The field sizes measured varied from 2cm×2cm to 40cm×40cm. All measurements were taken at 100cm SSD with an A1SL ion chamber. Results: Our results showed a significant (about twofold) increase in the dose rate for all field sizes. Also, a steeper reduction in depth doses was observed when the flattening filter was removed. No decrease was observed in the out‐of‐field dose, but there was an increase of the dose in the build‐up region that can be attributed to the increase of contaminant electrons. Conclusion:. Our study revealed that, for recent radiotherapy techniques, removing the flattening filter could be beneficial since there is a an increased dose rate and higher output therefore less monitor are required for each fraction leading to reduced whole body dose equivalent. On the other hand, higher skin doses are observed per monitor unit delivered but the overall skin dose for a complete treatment is comparable to the one with flattening filter since less monitor unit are required per fraction.

UR - http://www.scopus.com/inward/record.url?scp=85024816360&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024816360&partnerID=8YFLogxK

U2 - 10.1118/1.2961996

DO - 10.1118/1.2961996

M3 - Article

AN - SCOPUS:85024816360

VL - 35

SP - 2781

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -