SU‐GG‐T‐101

Validation of Collapse Cone Convolution Superposition Algorithm Using Ion Chamber and Field Measurements

O. Calvo, C. Buckey, Sotirios Stathakis, A. Gutiérrez, Nikos Papanikolaou

Research output: Contribution to journalArticle

Abstract

Purpose: To validate the collapse cone convolution superposition (CCCS) algorithm using ion chamber and field measurements for small lung lesions subject to electronic disequilibrium. Method and Materials: IMRT plans for several lung patients were created using Pinnacle 8.0 planning system. The lung lesions measured less than 3 cm diameter and less than 27 cm3 and were treated in our institution with SBRT. For each patient seven optimized plans were produced by changing the minimum segment size in the IMRT optimization parameters. The minimum segment size varied from 0.25cm2 to 6cm2. The optimized intensity maps for each plan were then used to calculate the dose distributions using the CCCS algorithm. For each clinical plan, a corresponding quality assurance (QA) plan was created. The intensity maps for each of the QA plans was calculated in a plexiglass phantom where there is an insert for an ion chamber (PTW 0.125 cc) and a film (Kodak EDR2). The ion chamber measurements were then converted to dose and the films were scanned using a VIDAR scanner and analyzed using the RIT software. The dose calculated from Pinnacle for the ion chamber point and the planar dose at the location of the film were then exported and compared against measurements. Results There was good agreement (0–5%) between the calculated values and the measurements. The agreement was independent of the segment size used during IMRT optimization. The plans with small segment sizes had more monitor units which resulted in higher doses to the lung. Conclusion: The CCCS can accurately predict the dose delivered to the target. Films and ion chamber measurements can be used for evaluating the delivered plan with accuracy of 5%.

Original languageEnglish (US)
Pages (from-to)2749
Number of pages1
JournalMedical Physics
Volume35
Issue number6
DOIs
StatePublished - 2008

Fingerprint

Ions
Lung
Polymethyl Methacrylate
Software

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

SU‐GG‐T‐101 : Validation of Collapse Cone Convolution Superposition Algorithm Using Ion Chamber and Field Measurements. / Calvo, O.; Buckey, C.; Stathakis, Sotirios; Gutiérrez, A.; Papanikolaou, Nikos.

In: Medical Physics, Vol. 35, No. 6, 2008, p. 2749.

Research output: Contribution to journalArticle

@article{96d3f71ce795400f913c7e9dcbc5a958,
title = "SU‐GG‐T‐101: Validation of Collapse Cone Convolution Superposition Algorithm Using Ion Chamber and Field Measurements",
abstract = "Purpose: To validate the collapse cone convolution superposition (CCCS) algorithm using ion chamber and field measurements for small lung lesions subject to electronic disequilibrium. Method and Materials: IMRT plans for several lung patients were created using Pinnacle 8.0 planning system. The lung lesions measured less than 3 cm diameter and less than 27 cm3 and were treated in our institution with SBRT. For each patient seven optimized plans were produced by changing the minimum segment size in the IMRT optimization parameters. The minimum segment size varied from 0.25cm2 to 6cm2. The optimized intensity maps for each plan were then used to calculate the dose distributions using the CCCS algorithm. For each clinical plan, a corresponding quality assurance (QA) plan was created. The intensity maps for each of the QA plans was calculated in a plexiglass phantom where there is an insert for an ion chamber (PTW 0.125 cc) and a film (Kodak EDR2). The ion chamber measurements were then converted to dose and the films were scanned using a VIDAR scanner and analyzed using the RIT software. The dose calculated from Pinnacle for the ion chamber point and the planar dose at the location of the film were then exported and compared against measurements. Results There was good agreement (0–5{\%}) between the calculated values and the measurements. The agreement was independent of the segment size used during IMRT optimization. The plans with small segment sizes had more monitor units which resulted in higher doses to the lung. Conclusion: The CCCS can accurately predict the dose delivered to the target. Films and ion chamber measurements can be used for evaluating the delivered plan with accuracy of 5{\%}.",
author = "O. Calvo and C. Buckey and Sotirios Stathakis and A. Guti{\'e}rrez and Nikos Papanikolaou",
year = "2008",
doi = "10.1118/1.2961853",
language = "English (US)",
volume = "35",
pages = "2749",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐GG‐T‐101

T2 - Validation of Collapse Cone Convolution Superposition Algorithm Using Ion Chamber and Field Measurements

AU - Calvo, O.

AU - Buckey, C.

AU - Stathakis, Sotirios

AU - Gutiérrez, A.

AU - Papanikolaou, Nikos

PY - 2008

Y1 - 2008

N2 - Purpose: To validate the collapse cone convolution superposition (CCCS) algorithm using ion chamber and field measurements for small lung lesions subject to electronic disequilibrium. Method and Materials: IMRT plans for several lung patients were created using Pinnacle 8.0 planning system. The lung lesions measured less than 3 cm diameter and less than 27 cm3 and were treated in our institution with SBRT. For each patient seven optimized plans were produced by changing the minimum segment size in the IMRT optimization parameters. The minimum segment size varied from 0.25cm2 to 6cm2. The optimized intensity maps for each plan were then used to calculate the dose distributions using the CCCS algorithm. For each clinical plan, a corresponding quality assurance (QA) plan was created. The intensity maps for each of the QA plans was calculated in a plexiglass phantom where there is an insert for an ion chamber (PTW 0.125 cc) and a film (Kodak EDR2). The ion chamber measurements were then converted to dose and the films were scanned using a VIDAR scanner and analyzed using the RIT software. The dose calculated from Pinnacle for the ion chamber point and the planar dose at the location of the film were then exported and compared against measurements. Results There was good agreement (0–5%) between the calculated values and the measurements. The agreement was independent of the segment size used during IMRT optimization. The plans with small segment sizes had more monitor units which resulted in higher doses to the lung. Conclusion: The CCCS can accurately predict the dose delivered to the target. Films and ion chamber measurements can be used for evaluating the delivered plan with accuracy of 5%.

AB - Purpose: To validate the collapse cone convolution superposition (CCCS) algorithm using ion chamber and field measurements for small lung lesions subject to electronic disequilibrium. Method and Materials: IMRT plans for several lung patients were created using Pinnacle 8.0 planning system. The lung lesions measured less than 3 cm diameter and less than 27 cm3 and were treated in our institution with SBRT. For each patient seven optimized plans were produced by changing the minimum segment size in the IMRT optimization parameters. The minimum segment size varied from 0.25cm2 to 6cm2. The optimized intensity maps for each plan were then used to calculate the dose distributions using the CCCS algorithm. For each clinical plan, a corresponding quality assurance (QA) plan was created. The intensity maps for each of the QA plans was calculated in a plexiglass phantom where there is an insert for an ion chamber (PTW 0.125 cc) and a film (Kodak EDR2). The ion chamber measurements were then converted to dose and the films were scanned using a VIDAR scanner and analyzed using the RIT software. The dose calculated from Pinnacle for the ion chamber point and the planar dose at the location of the film were then exported and compared against measurements. Results There was good agreement (0–5%) between the calculated values and the measurements. The agreement was independent of the segment size used during IMRT optimization. The plans with small segment sizes had more monitor units which resulted in higher doses to the lung. Conclusion: The CCCS can accurately predict the dose delivered to the target. Films and ion chamber measurements can be used for evaluating the delivered plan with accuracy of 5%.

UR - http://www.scopus.com/inward/record.url?scp=85024784824&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024784824&partnerID=8YFLogxK

U2 - 10.1118/1.2961853

DO - 10.1118/1.2961853

M3 - Article

VL - 35

SP - 2749

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -