MO‐F‐BRA‐04

Neutron Dose Equivalent at the Maze Entrance of GRID Therapy

X. Wang, M. Charlton, C. Esquivel, Sotirios Stathakis, Nikos Papanikolaou

Research output: Contribution to journalArticle

Abstract

Purpose: To evaluate the neutron and photon dose equivalent (Hn,D and HG) at the outer maze entrance with a grid block on a Varian Clinac 23EX accelerator; and to evaluate the neutron activation on the surface of the grid. Method and Materials:A Varian Clinac 23EX accelerator was used to produce 18 MV photons with a 7.62 cm brass grid. The neutron dose equivalent Hn,D was measured using an Andersson‐Braun neutron Rem meter, and the photon dose equivalent HG was measured using a Geiger Müller gamma‐ray survey meter at the outer maze entrance. The neutron activation dose on the surface of the grid was measured by the Geiger Müller survey meter after the irradiation. Results: With the gantry head tilted close to the inner maze entrance and with the jaws closed, the neutron dose equivalent reached its maximum, which is the same as the non‐grid beam. The difference between the grid and non‐grid measurement results was within the measurement uncertainty for different gantry angles. The measurement taken at 5 min after the irradiation shows the exposure rate at the surface of the grid can reach up to 360 mR/hr. The neutron activation activities have an average half life about 10 min. Conclusions: This work indicates that the neutron production in GRID therapy does not change from a conventional external beam therapy. However, the users of a grid should be aware of the possible high dose to the radiation worker from the neutron activation on the surface of the grid.

Original languageEnglish (US)
Pages (from-to)3721
Number of pages1
JournalMedical Physics
Volume38
Issue number6
DOIs
StatePublished - 2011

Fingerprint

Neutrons
Photons
Therapeutics
Jaw
Uncertainty
Half-Life
Head
Radiation

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

MO‐F‐BRA‐04 : Neutron Dose Equivalent at the Maze Entrance of GRID Therapy. / Wang, X.; Charlton, M.; Esquivel, C.; Stathakis, Sotirios; Papanikolaou, Nikos.

In: Medical Physics, Vol. 38, No. 6, 2011, p. 3721.

Research output: Contribution to journalArticle

Wang, X. ; Charlton, M. ; Esquivel, C. ; Stathakis, Sotirios ; Papanikolaou, Nikos. / MO‐F‐BRA‐04 : Neutron Dose Equivalent at the Maze Entrance of GRID Therapy. In: Medical Physics. 2011 ; Vol. 38, No. 6. pp. 3721.
@article{b9b93f652eed4784ba46555ce8b71bf7,
title = "MO‐F‐BRA‐04: Neutron Dose Equivalent at the Maze Entrance of GRID Therapy",
abstract = "Purpose: To evaluate the neutron and photon dose equivalent (Hn,D and HG) at the outer maze entrance with a grid block on a Varian Clinac 23EX accelerator; and to evaluate the neutron activation on the surface of the grid. Method and Materials:A Varian Clinac 23EX accelerator was used to produce 18 MV photons with a 7.62 cm brass grid. The neutron dose equivalent Hn,D was measured using an Andersson‐Braun neutron Rem meter, and the photon dose equivalent HG was measured using a Geiger M{\"u}ller gamma‐ray survey meter at the outer maze entrance. The neutron activation dose on the surface of the grid was measured by the Geiger M{\"u}ller survey meter after the irradiation. Results: With the gantry head tilted close to the inner maze entrance and with the jaws closed, the neutron dose equivalent reached its maximum, which is the same as the non‐grid beam. The difference between the grid and non‐grid measurement results was within the measurement uncertainty for different gantry angles. The measurement taken at 5 min after the irradiation shows the exposure rate at the surface of the grid can reach up to 360 mR/hr. The neutron activation activities have an average half life about 10 min. Conclusions: This work indicates that the neutron production in GRID therapy does not change from a conventional external beam therapy. However, the users of a grid should be aware of the possible high dose to the radiation worker from the neutron activation on the surface of the grid.",
author = "X. Wang and M. Charlton and C. Esquivel and Sotirios Stathakis and Nikos Papanikolaou",
year = "2011",
doi = "10.1118/1.3612999",
language = "English (US)",
volume = "38",
pages = "3721",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - MO‐F‐BRA‐04

T2 - Neutron Dose Equivalent at the Maze Entrance of GRID Therapy

AU - Wang, X.

AU - Charlton, M.

AU - Esquivel, C.

AU - Stathakis, Sotirios

AU - Papanikolaou, Nikos

PY - 2011

Y1 - 2011

N2 - Purpose: To evaluate the neutron and photon dose equivalent (Hn,D and HG) at the outer maze entrance with a grid block on a Varian Clinac 23EX accelerator; and to evaluate the neutron activation on the surface of the grid. Method and Materials:A Varian Clinac 23EX accelerator was used to produce 18 MV photons with a 7.62 cm brass grid. The neutron dose equivalent Hn,D was measured using an Andersson‐Braun neutron Rem meter, and the photon dose equivalent HG was measured using a Geiger Müller gamma‐ray survey meter at the outer maze entrance. The neutron activation dose on the surface of the grid was measured by the Geiger Müller survey meter after the irradiation. Results: With the gantry head tilted close to the inner maze entrance and with the jaws closed, the neutron dose equivalent reached its maximum, which is the same as the non‐grid beam. The difference between the grid and non‐grid measurement results was within the measurement uncertainty for different gantry angles. The measurement taken at 5 min after the irradiation shows the exposure rate at the surface of the grid can reach up to 360 mR/hr. The neutron activation activities have an average half life about 10 min. Conclusions: This work indicates that the neutron production in GRID therapy does not change from a conventional external beam therapy. However, the users of a grid should be aware of the possible high dose to the radiation worker from the neutron activation on the surface of the grid.

AB - Purpose: To evaluate the neutron and photon dose equivalent (Hn,D and HG) at the outer maze entrance with a grid block on a Varian Clinac 23EX accelerator; and to evaluate the neutron activation on the surface of the grid. Method and Materials:A Varian Clinac 23EX accelerator was used to produce 18 MV photons with a 7.62 cm brass grid. The neutron dose equivalent Hn,D was measured using an Andersson‐Braun neutron Rem meter, and the photon dose equivalent HG was measured using a Geiger Müller gamma‐ray survey meter at the outer maze entrance. The neutron activation dose on the surface of the grid was measured by the Geiger Müller survey meter after the irradiation. Results: With the gantry head tilted close to the inner maze entrance and with the jaws closed, the neutron dose equivalent reached its maximum, which is the same as the non‐grid beam. The difference between the grid and non‐grid measurement results was within the measurement uncertainty for different gantry angles. The measurement taken at 5 min after the irradiation shows the exposure rate at the surface of the grid can reach up to 360 mR/hr. The neutron activation activities have an average half life about 10 min. Conclusions: This work indicates that the neutron production in GRID therapy does not change from a conventional external beam therapy. However, the users of a grid should be aware of the possible high dose to the radiation worker from the neutron activation on the surface of the grid.

UR - http://www.scopus.com/inward/record.url?scp=85024823392&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024823392&partnerID=8YFLogxK

U2 - 10.1118/1.3612999

DO - 10.1118/1.3612999

M3 - Article

VL - 38

SP - 3721

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -