To evaluate the influence of stent design on endothelialization of the stented surface, we placed trapezoidal objects of variable thickness on a confluent culture of endothelial cells and subjected the assembly to flow and shear conditions similar to those found in arteries. After 24 h, we measured and analyzed the area on top of the objects covered by cells and the maximum migration distance from the borders. In addition, we evaluated areas devoid of cells surrounding the objects, which developed after exposure of the assembly to flow. The cell-covered area and migration distance significantly decreased on objects 75 μm thick, and it was nonexistent on objects 250 μm thick. Areas devoid of cells or gaps were largest adjacent to the downflow side of the object, disposed transversely to flow. Cell gaps were smallest along the side aligned with flow. In conclusion, endothelial cell coverage may be impaired by stent wall thickness larger than 75 μm. It is likely that this impairment is related to flow disturbances impairing cell attachment.

Original languageEnglish (US)
Pages (from-to)143-151
Number of pages9
JournalJournal of Long-Term Effects of Medical Implants
Volume10
Issue number1-2
StatePublished - 2000

Fingerprint

Stents
Endothelial cells
Topography
Cell culture
Endothelial Cells
Cell Movement
Arteries

Keywords

  • Endothelium
  • Flow disturbances
  • Stents

ASJC Scopus subject areas

  • Biomedical Engineering
  • Medicine (miscellaneous)
  • Orthopedics and Sports Medicine
  • Pathology and Forensic Medicine

Cite this

Influence of topography on endothelialization of stents : Clues for new designs. / Simon, Cristina; Palmaz, Julio C; Sprague, Eugene A.

In: Journal of Long-Term Effects of Medical Implants, Vol. 10, No. 1-2, 2000, p. 143-151.

Research output: Contribution to journalArticle

Simon, Cristina ; Palmaz, Julio C ; Sprague, Eugene A. / Influence of topography on endothelialization of stents : Clues for new designs. In: Journal of Long-Term Effects of Medical Implants. 2000 ; Vol. 10, No. 1-2. pp. 143-151.
@article{777ce65005b14a889cd182dff81b7a5b,
title = "Influence of topography on endothelialization of stents: Clues for new designs",
abstract = "To evaluate the influence of stent design on endothelialization of the stented surface, we placed trapezoidal objects of variable thickness on a confluent culture of endothelial cells and subjected the assembly to flow and shear conditions similar to those found in arteries. After 24 h, we measured and analyzed the area on top of the objects covered by cells and the maximum migration distance from the borders. In addition, we evaluated areas devoid of cells surrounding the objects, which developed after exposure of the assembly to flow. The cell-covered area and migration distance significantly decreased on objects 75 μm thick, and it was nonexistent on objects 250 μm thick. Areas devoid of cells or gaps were largest adjacent to the downflow side of the object, disposed transversely to flow. Cell gaps were smallest along the side aligned with flow. In conclusion, endothelial cell coverage may be impaired by stent wall thickness larger than 75 μm. It is likely that this impairment is related to flow disturbances impairing cell attachment.",
keywords = "Endothelium, Flow disturbances, Stents",
author = "Cristina Simon and Palmaz, {Julio C} and Sprague, {Eugene A}",
year = "2000",
language = "English (US)",
volume = "10",
pages = "143--151",
journal = "Journal of Long-Term Effects of Medical Implants",
issn = "1050-6934",
publisher = "Begell House Inc.",
number = "1-2",

}

TY - JOUR

T1 - Influence of topography on endothelialization of stents

T2 - Clues for new designs

AU - Simon, Cristina

AU - Palmaz, Julio C

AU - Sprague, Eugene A

PY - 2000

Y1 - 2000

N2 - To evaluate the influence of stent design on endothelialization of the stented surface, we placed trapezoidal objects of variable thickness on a confluent culture of endothelial cells and subjected the assembly to flow and shear conditions similar to those found in arteries. After 24 h, we measured and analyzed the area on top of the objects covered by cells and the maximum migration distance from the borders. In addition, we evaluated areas devoid of cells surrounding the objects, which developed after exposure of the assembly to flow. The cell-covered area and migration distance significantly decreased on objects 75 μm thick, and it was nonexistent on objects 250 μm thick. Areas devoid of cells or gaps were largest adjacent to the downflow side of the object, disposed transversely to flow. Cell gaps were smallest along the side aligned with flow. In conclusion, endothelial cell coverage may be impaired by stent wall thickness larger than 75 μm. It is likely that this impairment is related to flow disturbances impairing cell attachment.

AB - To evaluate the influence of stent design on endothelialization of the stented surface, we placed trapezoidal objects of variable thickness on a confluent culture of endothelial cells and subjected the assembly to flow and shear conditions similar to those found in arteries. After 24 h, we measured and analyzed the area on top of the objects covered by cells and the maximum migration distance from the borders. In addition, we evaluated areas devoid of cells surrounding the objects, which developed after exposure of the assembly to flow. The cell-covered area and migration distance significantly decreased on objects 75 μm thick, and it was nonexistent on objects 250 μm thick. Areas devoid of cells or gaps were largest adjacent to the downflow side of the object, disposed transversely to flow. Cell gaps were smallest along the side aligned with flow. In conclusion, endothelial cell coverage may be impaired by stent wall thickness larger than 75 μm. It is likely that this impairment is related to flow disturbances impairing cell attachment.

KW - Endothelium

KW - Flow disturbances

KW - Stents

UR - http://www.scopus.com/inward/record.url?scp=0034080259&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0034080259&partnerID=8YFLogxK

M3 - Article

C2 - 10947627

AN - SCOPUS:0034080259

VL - 10

SP - 143

EP - 151

JO - Journal of Long-Term Effects of Medical Implants

JF - Journal of Long-Term Effects of Medical Implants

SN - 1050-6934

IS - 1-2

ER -