Evaluation of the setup accuracy of a stereotactic radiotherapy head immobilization mask system using kV on-board imaging

Imad Ali, Jesse Tubbs, Kerry Hibbitts, Ozer Algan, Spencer Thompson, Terence Herman, Salahuddin Ahmad

Research output: Contribution to journalArticlepeer-review

17 Scopus citations

Abstract

The purpose of this study was to evaluate setup accuracy and quantify random and systematic errors of the BrainLAB stereotactic immobilization mask and localization system using kV on-board imaging. Nine patients were simulated and set up with the BrainLAB stereotactic head immobilization mask and localizer to be treated for brain lesions using single and hypofractions. Orthogonal pairs of projections were acquired using a kV on-board imager mounted on a Varian Trilogy machine. The kV projections were then registered with digitally-reconstructed radiographs (DRR) obtained from treatment planning. Shifts between the kV images and reference DRRs were calculated in the different directions: anterior-posterior (A-P), medial-lateral (R-L) and superior-inferior (S-I). If the shifts were larger than 2mm in any direction, the patient was reset within the immobilization mask until satisfying setup accuracy based on image guidance has been achieved. Shifts as large as 4.5 mm, 5.0 mm, 8.0 mm in the A-P, R-L and S-I directions, respectively, were measured from image registration of kV projections and DRRs. These shifts represent offsets between the treatment and simulation setup using immobilization mask. The mean offsets of 0.1 mm, 0.7 mm, and -1.6 mm represent systematic errors of the BrainLAB localizer in the A-P, R-L and S-I directions, respectively. The mean of the radial shifts is about 1.7 mm. The standard deviations of the shifts were 2.2 mm, 2.0 mm, and 2.6 mm in A-P, R-L and S-I directions, respectively, which represent random patient setup errors with the BrainLAB mask. The BrainLAB mask provides a noninvasive, practical and flexible immobilization system that keeps the patients in place during treatment. Relying on this system for patient setup might be associated with significant setup errors. Image guidance with the kV on-board imager provides an independent verification technique to ensure accuracy of patient setup. Since the patient may relax or move during treatment, uncontrolled and undetected setup errors may be produced with patients that are not well-immobilized. Therefore, the combination of stereotactic immobilization and image guidance achieves more controlled and accurate patient setup within 2mm in A-P, R-L and S-I directions.

Original languageEnglish (US)
Pages (from-to)26-37
Number of pages12
JournalJournal of Applied Clinical Medical Physics
Volume11
Issue number3
DOIs
StatePublished - 2010
Externally publishedYes

Keywords

  • BrainLAB
  • On-board imager
  • Stereotactic radiation therapy
  • Systematic and random errors
  • Thermoplastic immobilization mask

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Evaluation of the setup accuracy of a stereotactic radiotherapy head immobilization mask system using kV on-board imaging'. Together they form a unique fingerprint.

Cite this