Effects of fiberoptic bronchoscopy during mechanical ventilation in a lung model

Research output: Contribution to journalArticlepeer-review

55 Scopus citations

Abstract

Study objectives: To evaluate the effects of fiberoptic bronchoscopy (FOB) on delivered volumes and pressures during mechanical ventilation, utilizing a lung model. Design: Bench study. Setting: Laboratory. Materials and methods: Using varying-sized endotracheal tubes (ETTs), we ventilated a lung model at two levels of compliance utilizing different modes and parameters of ventilation. After establishing baseline measurements, the bronchoscope was inserted and measurements repeated. Measurements and results: During controlled mechanical ventilation (CMV) with a preset high-pressure limit (HPL), tidal volumes (VTs) were reduced from 700 mL to 0 to 500 mL following insertion of the bronchoscope. Increasing the HPL to 120 cm H2O resulted in a VT of 40 to 680 mL. Changing from square to decelerating flow waveform resulted in no consistent difference in VT. Auto-positive end-expiratory pressure (auto-PEEP) of 0 to 41 cm H2O was present under most conditions. Higher rates and lower peak inspiratory flows were associated with higher levels of auto-PEEP. In the pressure-control (PC) mode, using a preset inspiratory pressure level (IP), VT was reduced from 700 mL to 40 to 280 mL following insertion of the bronchoscope. Maximum IP (100 cm H2O) increased VT to 260 to 700 mL. Auto-PEEP was less in the PC mode. Conclusions: Extreme care must be taken when bronchoscopy is performed on a patient receiving mechanical ventilation. Extremely low VT and significant auto-PEEP may develop unless flow, respiratory rate, mode, and ETT size are carefully selected. The PC mode delivered more volume than did the CMV mode. When performing FOB during mechanical ventilation, the inside diameter of the ETT should be ≥ 2.0-mm larger than the outside diameter of the bronchoscope to maintain volume delivery and minimize the development of auto-PEEP.

Original languageEnglish (US)
Pages (from-to)824-831
Number of pages8
JournalChest
Volume118
Issue number3
DOIs
StatePublished - 2000

Keywords

  • Auto-positive end-expiratory pressure
  • Bronchoscopy
  • Mechanical ventilation
  • Peak inspiratory pressure

ASJC Scopus subject areas

  • Critical Care and Intensive Care Medicine
  • Cardiology and Cardiovascular Medicine
  • Pulmonary and Respiratory Medicine

Fingerprint

Dive into the research topics of 'Effects of fiberoptic bronchoscopy during mechanical ventilation in a lung model'. Together they form a unique fingerprint.

Cite this