Effects of CT Section Thickness and Reconstruction Kernel on Emphysema Quantification. Relationship to the Magnitude of the CT Emphysema Index

David S. Gierada, Andrew J. Bierhals, Cliff K. Choong, Seth T. Bartel, Jon H. Ritter, Nitin A Das, Cheng Hong, Thomas K. Pilgram, Kyongtae T. Bae, Bruce R. Whiting, Jason C. Woods, James C. Hogg, Barbara A. Lutey, Richard J. Battafarano, Joel D. Cooper, Bryan F. Meyers, G. Alexander Patterson

Research output: Contribution to journalArticle

59 Citations (Scopus)

Abstract

Rationale and Objectives: Computed tomography (CT) section thickness and reconstruction kernel each influence CT measurements of emphysema. This study was performed to assess whether their effects are related to the magnitude of the measurement. Materials and Methods: Low-radiation-dose multidetector CT was performed in 21 subjects representing a wide range of emphysema severity. Images were reconstructed using 20 different combinations of section thickness and reconstruction kernel. Emphysema index values were determined as the percentage of lung pixels having attenuation lower than multiple thresholds ranging from -960 HU to -890 HU. The index values obtained from the different thickness-kernel combinations were compared by repeated measures analysis of variance and Bland-Altman plots of mean versus difference in all subjects, and correlated with quantitative histology (mean linear intercept, Lm) in a subset of resected lung specimens. Results: The effects of section thickness and reconstruction kernel on the emphysema index were significant (P < .001) and diminished as the index attenuation threshold was raised. The changes in index values from changing the thickness-kernel combination were largest for subjects with intermediate index values (10%-30%), and became progressively smaller for those with lower and higher index values. This pattern was consistent regardless of the thickness-kernel combinations compared and the HU threshold used. Correlations between the emphysema index values obtained with each thickness-kernel combination and Lm ranged from r = 0.55-0.68 (P = .007-.03). Conclusion: The effects of CT section thickness and kernel on emphysema index values varied systematically with the magnitude of the emphysema index. All reconstruction techniques provided significant correlations with quantitative histology.

Original languageEnglish (US)
Pages (from-to)146-156
Number of pages11
JournalAcademic Radiology
Volume17
Issue number2
DOIs
StatePublished - Feb 2010
Externally publishedYes

Fingerprint

Emphysema
Tomography
Histology
Lung
Multidetector Computed Tomography
Analysis of Variance
Radiation

Keywords

  • computed tomography
  • Emphysema
  • histology

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Gierada, D. S., Bierhals, A. J., Choong, C. K., Bartel, S. T., Ritter, J. H., Das, N. A., ... Patterson, G. A. (2010). Effects of CT Section Thickness and Reconstruction Kernel on Emphysema Quantification. Relationship to the Magnitude of the CT Emphysema Index. Academic Radiology, 17(2), 146-156. https://doi.org/10.1016/j.acra.2009.08.007

Effects of CT Section Thickness and Reconstruction Kernel on Emphysema Quantification. Relationship to the Magnitude of the CT Emphysema Index. / Gierada, David S.; Bierhals, Andrew J.; Choong, Cliff K.; Bartel, Seth T.; Ritter, Jon H.; Das, Nitin A; Hong, Cheng; Pilgram, Thomas K.; Bae, Kyongtae T.; Whiting, Bruce R.; Woods, Jason C.; Hogg, James C.; Lutey, Barbara A.; Battafarano, Richard J.; Cooper, Joel D.; Meyers, Bryan F.; Patterson, G. Alexander.

In: Academic Radiology, Vol. 17, No. 2, 02.2010, p. 146-156.

Research output: Contribution to journalArticle

Gierada, DS, Bierhals, AJ, Choong, CK, Bartel, ST, Ritter, JH, Das, NA, Hong, C, Pilgram, TK, Bae, KT, Whiting, BR, Woods, JC, Hogg, JC, Lutey, BA, Battafarano, RJ, Cooper, JD, Meyers, BF & Patterson, GA 2010, 'Effects of CT Section Thickness and Reconstruction Kernel on Emphysema Quantification. Relationship to the Magnitude of the CT Emphysema Index', Academic Radiology, vol. 17, no. 2, pp. 146-156. https://doi.org/10.1016/j.acra.2009.08.007
Gierada, David S. ; Bierhals, Andrew J. ; Choong, Cliff K. ; Bartel, Seth T. ; Ritter, Jon H. ; Das, Nitin A ; Hong, Cheng ; Pilgram, Thomas K. ; Bae, Kyongtae T. ; Whiting, Bruce R. ; Woods, Jason C. ; Hogg, James C. ; Lutey, Barbara A. ; Battafarano, Richard J. ; Cooper, Joel D. ; Meyers, Bryan F. ; Patterson, G. Alexander. / Effects of CT Section Thickness and Reconstruction Kernel on Emphysema Quantification. Relationship to the Magnitude of the CT Emphysema Index. In: Academic Radiology. 2010 ; Vol. 17, No. 2. pp. 146-156.
@article{b15c6176548641b99c46c1a5bb33b81b,
title = "Effects of CT Section Thickness and Reconstruction Kernel on Emphysema Quantification. Relationship to the Magnitude of the CT Emphysema Index",
abstract = "Rationale and Objectives: Computed tomography (CT) section thickness and reconstruction kernel each influence CT measurements of emphysema. This study was performed to assess whether their effects are related to the magnitude of the measurement. Materials and Methods: Low-radiation-dose multidetector CT was performed in 21 subjects representing a wide range of emphysema severity. Images were reconstructed using 20 different combinations of section thickness and reconstruction kernel. Emphysema index values were determined as the percentage of lung pixels having attenuation lower than multiple thresholds ranging from -960 HU to -890 HU. The index values obtained from the different thickness-kernel combinations were compared by repeated measures analysis of variance and Bland-Altman plots of mean versus difference in all subjects, and correlated with quantitative histology (mean linear intercept, Lm) in a subset of resected lung specimens. Results: The effects of section thickness and reconstruction kernel on the emphysema index were significant (P < .001) and diminished as the index attenuation threshold was raised. The changes in index values from changing the thickness-kernel combination were largest for subjects with intermediate index values (10{\%}-30{\%}), and became progressively smaller for those with lower and higher index values. This pattern was consistent regardless of the thickness-kernel combinations compared and the HU threshold used. Correlations between the emphysema index values obtained with each thickness-kernel combination and Lm ranged from r = 0.55-0.68 (P = .007-.03). Conclusion: The effects of CT section thickness and kernel on emphysema index values varied systematically with the magnitude of the emphysema index. All reconstruction techniques provided significant correlations with quantitative histology.",
keywords = "computed tomography, Emphysema, histology",
author = "Gierada, {David S.} and Bierhals, {Andrew J.} and Choong, {Cliff K.} and Bartel, {Seth T.} and Ritter, {Jon H.} and Das, {Nitin A} and Cheng Hong and Pilgram, {Thomas K.} and Bae, {Kyongtae T.} and Whiting, {Bruce R.} and Woods, {Jason C.} and Hogg, {James C.} and Lutey, {Barbara A.} and Battafarano, {Richard J.} and Cooper, {Joel D.} and Meyers, {Bryan F.} and Patterson, {G. Alexander}",
year = "2010",
month = "2",
doi = "10.1016/j.acra.2009.08.007",
language = "English (US)",
volume = "17",
pages = "146--156",
journal = "Academic Radiology",
issn = "1076-6332",
publisher = "Elsevier USA",
number = "2",

}

TY - JOUR

T1 - Effects of CT Section Thickness and Reconstruction Kernel on Emphysema Quantification. Relationship to the Magnitude of the CT Emphysema Index

AU - Gierada, David S.

AU - Bierhals, Andrew J.

AU - Choong, Cliff K.

AU - Bartel, Seth T.

AU - Ritter, Jon H.

AU - Das, Nitin A

AU - Hong, Cheng

AU - Pilgram, Thomas K.

AU - Bae, Kyongtae T.

AU - Whiting, Bruce R.

AU - Woods, Jason C.

AU - Hogg, James C.

AU - Lutey, Barbara A.

AU - Battafarano, Richard J.

AU - Cooper, Joel D.

AU - Meyers, Bryan F.

AU - Patterson, G. Alexander

PY - 2010/2

Y1 - 2010/2

N2 - Rationale and Objectives: Computed tomography (CT) section thickness and reconstruction kernel each influence CT measurements of emphysema. This study was performed to assess whether their effects are related to the magnitude of the measurement. Materials and Methods: Low-radiation-dose multidetector CT was performed in 21 subjects representing a wide range of emphysema severity. Images were reconstructed using 20 different combinations of section thickness and reconstruction kernel. Emphysema index values were determined as the percentage of lung pixels having attenuation lower than multiple thresholds ranging from -960 HU to -890 HU. The index values obtained from the different thickness-kernel combinations were compared by repeated measures analysis of variance and Bland-Altman plots of mean versus difference in all subjects, and correlated with quantitative histology (mean linear intercept, Lm) in a subset of resected lung specimens. Results: The effects of section thickness and reconstruction kernel on the emphysema index were significant (P < .001) and diminished as the index attenuation threshold was raised. The changes in index values from changing the thickness-kernel combination were largest for subjects with intermediate index values (10%-30%), and became progressively smaller for those with lower and higher index values. This pattern was consistent regardless of the thickness-kernel combinations compared and the HU threshold used. Correlations between the emphysema index values obtained with each thickness-kernel combination and Lm ranged from r = 0.55-0.68 (P = .007-.03). Conclusion: The effects of CT section thickness and kernel on emphysema index values varied systematically with the magnitude of the emphysema index. All reconstruction techniques provided significant correlations with quantitative histology.

AB - Rationale and Objectives: Computed tomography (CT) section thickness and reconstruction kernel each influence CT measurements of emphysema. This study was performed to assess whether their effects are related to the magnitude of the measurement. Materials and Methods: Low-radiation-dose multidetector CT was performed in 21 subjects representing a wide range of emphysema severity. Images were reconstructed using 20 different combinations of section thickness and reconstruction kernel. Emphysema index values were determined as the percentage of lung pixels having attenuation lower than multiple thresholds ranging from -960 HU to -890 HU. The index values obtained from the different thickness-kernel combinations were compared by repeated measures analysis of variance and Bland-Altman plots of mean versus difference in all subjects, and correlated with quantitative histology (mean linear intercept, Lm) in a subset of resected lung specimens. Results: The effects of section thickness and reconstruction kernel on the emphysema index were significant (P < .001) and diminished as the index attenuation threshold was raised. The changes in index values from changing the thickness-kernel combination were largest for subjects with intermediate index values (10%-30%), and became progressively smaller for those with lower and higher index values. This pattern was consistent regardless of the thickness-kernel combinations compared and the HU threshold used. Correlations between the emphysema index values obtained with each thickness-kernel combination and Lm ranged from r = 0.55-0.68 (P = .007-.03). Conclusion: The effects of CT section thickness and kernel on emphysema index values varied systematically with the magnitude of the emphysema index. All reconstruction techniques provided significant correlations with quantitative histology.

KW - computed tomography

KW - Emphysema

KW - histology

UR - http://www.scopus.com/inward/record.url?scp=73049095694&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=73049095694&partnerID=8YFLogxK

U2 - 10.1016/j.acra.2009.08.007

DO - 10.1016/j.acra.2009.08.007

M3 - Article

C2 - 19931472

AN - SCOPUS:73049095694

VL - 17

SP - 146

EP - 156

JO - Academic Radiology

JF - Academic Radiology

SN - 1076-6332

IS - 2

ER -