Effect of electron contamination on in vivo dosimetry for lung block shielding during TBI

Ganesh Narayanasamy, Wilbert Cruz, Daniel L. Saenz, Sotirios Stathakis, Niko Papanikolaou, Neil Kirby

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

Our institution performs in vivo verification measurement for each of our total body irradiation (TBI) patients with optically stimulated luminescent dosimeters (OSLD). The lung block verification measurements were commonly higher than expected. The aim of this work is to understand this discrepancy and improve the accuracy of these lung block verification measurements. Initially, the thickness of the lung block was increased to provide adequate lung sparing. Further tests revealed the increase was due to electron contamination dose emanating from the lung block. The thickness of the bolus material covering the OSLD behind the lung block was increased to offset the electron contamination. In addition, the distance from the lung block to the dosimeter was evaluated for its effect on the OSLD reading and found to be clinically insignificant over the range of variability in our clinic. The results show that the improved TBI treatment technique provides for better accuracy of measured dose in vivo and consistency of patient setup.

Original languageEnglish (US)
Pages (from-to)486-491
Number of pages6
JournalJournal of Applied Clinical Medical Physics
Volume17
Issue number3
DOIs
StatePublished - 2016

Keywords

  • Dosimetry
  • Electron contamination
  • Lung block
  • OSLD
  • TBI

ASJC Scopus subject areas

  • Instrumentation
  • Radiation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Effect of electron contamination on in vivo dosimetry for lung block shielding during TBI'. Together they form a unique fingerprint.

Cite this