TY - GEN
T1 - Dosimetric comparison of volumetric modulated Arc therapy, step-and-shoot, and sliding window IMRT for prostate cancer
AU - Schnell, Erich
AU - Herman, Tania De La Fuente
AU - Young, Julie
AU - Hildebrand, Kim
AU - Algan, Ozer
AU - Syzek, Elizabeth
AU - Herman, Terence
AU - Ahmad, Salahuddin
PY - 2012
Y1 - 2012
N2 - This study aims to evaluate treatment plans generated by Step-and-Shoot (SS), Sliding Window (SW) and Volumetric Modulated Arc Therapy (VMAT) in order to assess the differences in dose volume histograms of planning target volume (PTV) and organs at risk (OAR), conformity indices, radiobiological evaluations, and plan quality for prostate cancer cases. Six prostate cancer patients treated in our center were selected for this retrospective study. Treatment plans were generated with Eclipse version 8.9 using 10 MV photon beams. For VMAT, Varian Rapid Arc with 1 or 2 arcs, and for SS and SW IMRT, 7-9 fields were used. Each plan had three PTVs with prescription doses of 81, 59.4, and 45 Gy to prostate, to prostate and lymph nodes, and to pelvis, respectively. Doses to PTV and OAR and the conformal indices (COIN) were compared among three techniques. The equivalent uniform dose (EUD), tumor control probability (TCP) and normal tissue complication probability (NTCP) were calculated and compared. The mean doses to the PTV prostate on average were 83 Gy and the percent differences of mean dose among all techniques were below 0.28. For bladder and rectum, the percent differences of mean dose among all techniques were below 2.2. The COIN did not favour any particular delivery method over the other. The TCP was higher with SS and SW for four patients and higher with VMAT for two patients. The NTCP for the rectum was the lowest with VMAT in five out of the six patients. The results show similar target coverage in general.
AB - This study aims to evaluate treatment plans generated by Step-and-Shoot (SS), Sliding Window (SW) and Volumetric Modulated Arc Therapy (VMAT) in order to assess the differences in dose volume histograms of planning target volume (PTV) and organs at risk (OAR), conformity indices, radiobiological evaluations, and plan quality for prostate cancer cases. Six prostate cancer patients treated in our center were selected for this retrospective study. Treatment plans were generated with Eclipse version 8.9 using 10 MV photon beams. For VMAT, Varian Rapid Arc with 1 or 2 arcs, and for SS and SW IMRT, 7-9 fields were used. Each plan had three PTVs with prescription doses of 81, 59.4, and 45 Gy to prostate, to prostate and lymph nodes, and to pelvis, respectively. Doses to PTV and OAR and the conformal indices (COIN) were compared among three techniques. The equivalent uniform dose (EUD), tumor control probability (TCP) and normal tissue complication probability (NTCP) were calculated and compared. The mean doses to the PTV prostate on average were 83 Gy and the percent differences of mean dose among all techniques were below 0.28. For bladder and rectum, the percent differences of mean dose among all techniques were below 2.2. The COIN did not favour any particular delivery method over the other. The TCP was higher with SS and SW for four patients and higher with VMAT for two patients. The NTCP for the rectum was the lowest with VMAT in five out of the six patients. The results show similar target coverage in general.
KW - Intensity Modulated Radiation Therapy
KW - Prostate Cancer
KW - Sliding Window
KW - Step and Shoot
KW - Volumetric Modulated Arc Therapy
UR - http://www.scopus.com/inward/record.url?scp=84873658059&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84873658059&partnerID=8YFLogxK
U2 - 10.1063/1.4764589
DO - 10.1063/1.4764589
M3 - Conference contribution
AN - SCOPUS:84873658059
SN - 9780735411067
T3 - AIP Conference Proceedings
SP - 23
EP - 26
BT - Medical Physics - Twelfth Mexican Symposium on Medical Physics
T2 - 12th Mexican Symposium on Medical Physics
Y2 - 16 March 2012 through 19 March 2012
ER -