Dosimetric aspects of inverse-planned modulated-arc total-body irradiation

Mareike Held, Neil Kirby, Olivier Morin, Jean Pouliot

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

Purpose: To develop optimal beam parameters and to verify the dosimetric aspects of the recently developed modulated-arc total-body irradiation (MATBI) technique, which delivers an inverse-planned dose to the entire body using gantry rotation. Methods: The patient is positioned prone and supine underneath the gantry at about 2 m source-to-surface distance (SSD). Then, up to 28 beams irradiate the patient from different gantry angles. Based on full-body computed-tomography (CT) images of the patient, the weight of each beam is optimized, using inverse planning, to create a uniform body dose. This study investigates how to best simulate patients and the ideal beam setup parameters, such as field size, number of beams, and beam geometry, for treatment time and dose homogeneity. In addition, three anthropomorphic water phantoms were constructed and utilized to verify the accuracy of dose delivery, with both diode array and ion chamber measurements. Furthermore, to improve the accuracy of the new technique, a beam model is created specifically for the extended-SSD positioning for MATBI. Results: Low dose CT scans can be utilized for dose calculations without affecting the accuracy. The largest field size of 40 × 40 cm2 was found to deliver the most uniform dose in the least amount of time. Moreover, a higher number of beams improves dose homogeneity. The average dose discrepancy between ion chamber measurements and extended-SSD beam model calculations was 1.2, with the largest discrepancy being 3.2. This average dose discrepancy was 1.4 with the standard beam model for delivery at isocenter. Conclusions: The optimum beam setup parameters, regarding dose uniformity and treatment duration, are laid out for modulated-arc TBI. In addition, the presented dose measurements show that these treatments can be delivered accurately. These measurements also indicated that a new beam model did not significantly improve the accuracy of dose calculations. The optimum beam setup parameters along with the measurements performed to ensure accurate dose delivery serve as a useful guide for the clinical implementation of MATBI.

Original languageEnglish (US)
Pages (from-to)5263-5271
Number of pages9
JournalMedical Physics
Volume39
Issue number8
DOIs
StatePublished - Jan 1 2012
Externally publishedYes

Fingerprint

Whole-Body Irradiation
Tomography
Ions
Therapeutics
Weights and Measures
Water

Keywords

  • full-body dose distribution
  • inverse planning
  • TBI arc treatment
  • total-body irradiation (TBI)
  • treatment techniques

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Dosimetric aspects of inverse-planned modulated-arc total-body irradiation. / Held, Mareike; Kirby, Neil; Morin, Olivier; Pouliot, Jean.

In: Medical Physics, Vol. 39, No. 8, 01.01.2012, p. 5263-5271.

Research output: Contribution to journalArticle

Held, Mareike ; Kirby, Neil ; Morin, Olivier ; Pouliot, Jean. / Dosimetric aspects of inverse-planned modulated-arc total-body irradiation. In: Medical Physics. 2012 ; Vol. 39, No. 8. pp. 5263-5271.
@article{ebd5955919094da4acb0688b76414edb,
title = "Dosimetric aspects of inverse-planned modulated-arc total-body irradiation",
abstract = "Purpose: To develop optimal beam parameters and to verify the dosimetric aspects of the recently developed modulated-arc total-body irradiation (MATBI) technique, which delivers an inverse-planned dose to the entire body using gantry rotation. Methods: The patient is positioned prone and supine underneath the gantry at about 2 m source-to-surface distance (SSD). Then, up to 28 beams irradiate the patient from different gantry angles. Based on full-body computed-tomography (CT) images of the patient, the weight of each beam is optimized, using inverse planning, to create a uniform body dose. This study investigates how to best simulate patients and the ideal beam setup parameters, such as field size, number of beams, and beam geometry, for treatment time and dose homogeneity. In addition, three anthropomorphic water phantoms were constructed and utilized to verify the accuracy of dose delivery, with both diode array and ion chamber measurements. Furthermore, to improve the accuracy of the new technique, a beam model is created specifically for the extended-SSD positioning for MATBI. Results: Low dose CT scans can be utilized for dose calculations without affecting the accuracy. The largest field size of 40 × 40 cm2 was found to deliver the most uniform dose in the least amount of time. Moreover, a higher number of beams improves dose homogeneity. The average dose discrepancy between ion chamber measurements and extended-SSD beam model calculations was 1.2, with the largest discrepancy being 3.2. This average dose discrepancy was 1.4 with the standard beam model for delivery at isocenter. Conclusions: The optimum beam setup parameters, regarding dose uniformity and treatment duration, are laid out for modulated-arc TBI. In addition, the presented dose measurements show that these treatments can be delivered accurately. These measurements also indicated that a new beam model did not significantly improve the accuracy of dose calculations. The optimum beam setup parameters along with the measurements performed to ensure accurate dose delivery serve as a useful guide for the clinical implementation of MATBI.",
keywords = "full-body dose distribution, inverse planning, TBI arc treatment, total-body irradiation (TBI), treatment techniques",
author = "Mareike Held and Neil Kirby and Olivier Morin and Jean Pouliot",
year = "2012",
month = "1",
day = "1",
doi = "10.1118/1.4739250",
language = "English (US)",
volume = "39",
pages = "5263--5271",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "8",

}

TY - JOUR

T1 - Dosimetric aspects of inverse-planned modulated-arc total-body irradiation

AU - Held, Mareike

AU - Kirby, Neil

AU - Morin, Olivier

AU - Pouliot, Jean

PY - 2012/1/1

Y1 - 2012/1/1

N2 - Purpose: To develop optimal beam parameters and to verify the dosimetric aspects of the recently developed modulated-arc total-body irradiation (MATBI) technique, which delivers an inverse-planned dose to the entire body using gantry rotation. Methods: The patient is positioned prone and supine underneath the gantry at about 2 m source-to-surface distance (SSD). Then, up to 28 beams irradiate the patient from different gantry angles. Based on full-body computed-tomography (CT) images of the patient, the weight of each beam is optimized, using inverse planning, to create a uniform body dose. This study investigates how to best simulate patients and the ideal beam setup parameters, such as field size, number of beams, and beam geometry, for treatment time and dose homogeneity. In addition, three anthropomorphic water phantoms were constructed and utilized to verify the accuracy of dose delivery, with both diode array and ion chamber measurements. Furthermore, to improve the accuracy of the new technique, a beam model is created specifically for the extended-SSD positioning for MATBI. Results: Low dose CT scans can be utilized for dose calculations without affecting the accuracy. The largest field size of 40 × 40 cm2 was found to deliver the most uniform dose in the least amount of time. Moreover, a higher number of beams improves dose homogeneity. The average dose discrepancy between ion chamber measurements and extended-SSD beam model calculations was 1.2, with the largest discrepancy being 3.2. This average dose discrepancy was 1.4 with the standard beam model for delivery at isocenter. Conclusions: The optimum beam setup parameters, regarding dose uniformity and treatment duration, are laid out for modulated-arc TBI. In addition, the presented dose measurements show that these treatments can be delivered accurately. These measurements also indicated that a new beam model did not significantly improve the accuracy of dose calculations. The optimum beam setup parameters along with the measurements performed to ensure accurate dose delivery serve as a useful guide for the clinical implementation of MATBI.

AB - Purpose: To develop optimal beam parameters and to verify the dosimetric aspects of the recently developed modulated-arc total-body irradiation (MATBI) technique, which delivers an inverse-planned dose to the entire body using gantry rotation. Methods: The patient is positioned prone and supine underneath the gantry at about 2 m source-to-surface distance (SSD). Then, up to 28 beams irradiate the patient from different gantry angles. Based on full-body computed-tomography (CT) images of the patient, the weight of each beam is optimized, using inverse planning, to create a uniform body dose. This study investigates how to best simulate patients and the ideal beam setup parameters, such as field size, number of beams, and beam geometry, for treatment time and dose homogeneity. In addition, three anthropomorphic water phantoms were constructed and utilized to verify the accuracy of dose delivery, with both diode array and ion chamber measurements. Furthermore, to improve the accuracy of the new technique, a beam model is created specifically for the extended-SSD positioning for MATBI. Results: Low dose CT scans can be utilized for dose calculations without affecting the accuracy. The largest field size of 40 × 40 cm2 was found to deliver the most uniform dose in the least amount of time. Moreover, a higher number of beams improves dose homogeneity. The average dose discrepancy between ion chamber measurements and extended-SSD beam model calculations was 1.2, with the largest discrepancy being 3.2. This average dose discrepancy was 1.4 with the standard beam model for delivery at isocenter. Conclusions: The optimum beam setup parameters, regarding dose uniformity and treatment duration, are laid out for modulated-arc TBI. In addition, the presented dose measurements show that these treatments can be delivered accurately. These measurements also indicated that a new beam model did not significantly improve the accuracy of dose calculations. The optimum beam setup parameters along with the measurements performed to ensure accurate dose delivery serve as a useful guide for the clinical implementation of MATBI.

KW - full-body dose distribution

KW - inverse planning

KW - TBI arc treatment

KW - total-body irradiation (TBI)

KW - treatment techniques

UR - http://www.scopus.com/inward/record.url?scp=84864682262&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84864682262&partnerID=8YFLogxK

U2 - 10.1118/1.4739250

DO - 10.1118/1.4739250

M3 - Article

C2 - 22894451

AN - SCOPUS:84864682262

VL - 39

SP - 5263

EP - 5271

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 8

ER -