Comparison of the helical tomotherapy and MLC-based IMRT radiation modalities in treating brain and cranio-spinal tumors

Panayiotis Mavroidis, Brigida Costa Ferreira, Chengyu Shi, Miltiadis G. Delichas, Bengt K. Lind, Nikos Papanikolaou

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

The investigation of the clinical efficacy and effectiveness of Intensity Modulated Radiotherapy (IMRT) using Multileaf Collimators (MLC) and Helical Tomotherapy (HT) has been an issue of increasing interest over the past few years. In order to assess the suitability of a treatment plan, dosimetric criteria such as dose-volume histograms (DVH), maximum, minimum, mean, and standard deviation of the dose distribution are typically used. Nevertheless, the radiobiological parameters of the different tumors and normal tissues are often not taken into account. The use of the biologically effective uniform dose (D=) together with the complication-free tumor control probability (P+) were applied to evaluate the two radiation modalities. Two different clinical cases of brain and cranio-spinal axis cancers have been investigated by developing a linac MLC-based step-and-shoot IMRT plan and a Helical Tomotherapy plan. The treatment plans of the MLC-based IMRT were developed on the Philips treatment planning station using the Pinnacle 7.6 software release while the dedicated Tomotherapy treatment planning station was used for the HT plan. With the use of the P+ index and the D = concept as the common prescription point, the different treatment plans were compared based on radiobiological measures. The tissue response probabilities were plotted against D= for a range of prescription doses. The applied plan evaluation method shows that in the brain cancer, the HT treatment gives slightly better results than the MLC-based IMRT in terms of optimum expected clinical outcome (P+ of 66.1% and 63.5% for a D = to the PTV of 63.0 Gy and 62.0 Gy, respectively). In the cranio-spinal axis cancer, the HT plan is significantly better compared to the MLC-based IMRT plan over the clinically useful dose prescription range (P + of 84.1% and 28.3% for a D= to the PTV of 50.6 Gy and 44.0 Gy, respectively). If a higher than 5% risk for complications could be allowed, the complication-free tumor control could be increased by almost 30% compared to the initial dose prescription. In comparison to MLC based-IMRT, HT can better encompass the often large PTV while minimizing the volume of the OARs receiving high dose. A radiobiological treatment plan evaluation can provide a closer association of the delivered treatment with the clinical outcome by taking into account the dose-response relations of the irradiated tumors and normal tissues. The use of P - D= diagrams can complement the traditional tools of evaluation such as DVHs, in order to compare and effectively evaluate different treatment plans.

Original languageEnglish (US)
Pages (from-to)3-14
Number of pages12
JournalTechnology in Cancer Research and Treatment
Volume8
Issue number1
StatePublished - Feb 2009

Fingerprint

Intensity-Modulated Radiotherapy
Radiation
Brain
Neoplasms
Prescriptions
Therapeutics
Brain Neoplasms
Software

Keywords

  • Biologically effective uniform dose
  • Complication-free tumor control
  • Dose-response relations
  • Helical Tomotherapy
  • IMRT
  • Radiobiological evaluation

ASJC Scopus subject areas

  • Cancer Research
  • Oncology

Cite this

Comparison of the helical tomotherapy and MLC-based IMRT radiation modalities in treating brain and cranio-spinal tumors. / Mavroidis, Panayiotis; Ferreira, Brigida Costa; Shi, Chengyu; Delichas, Miltiadis G.; Lind, Bengt K.; Papanikolaou, Nikos.

In: Technology in Cancer Research and Treatment, Vol. 8, No. 1, 02.2009, p. 3-14.

Research output: Contribution to journalArticle

Mavroidis, Panayiotis ; Ferreira, Brigida Costa ; Shi, Chengyu ; Delichas, Miltiadis G. ; Lind, Bengt K. ; Papanikolaou, Nikos. / Comparison of the helical tomotherapy and MLC-based IMRT radiation modalities in treating brain and cranio-spinal tumors. In: Technology in Cancer Research and Treatment. 2009 ; Vol. 8, No. 1. pp. 3-14.
@article{4d1173ea25f14171827d5b24b61f9165,
title = "Comparison of the helical tomotherapy and MLC-based IMRT radiation modalities in treating brain and cranio-spinal tumors",
abstract = "The investigation of the clinical efficacy and effectiveness of Intensity Modulated Radiotherapy (IMRT) using Multileaf Collimators (MLC) and Helical Tomotherapy (HT) has been an issue of increasing interest over the past few years. In order to assess the suitability of a treatment plan, dosimetric criteria such as dose-volume histograms (DVH), maximum, minimum, mean, and standard deviation of the dose distribution are typically used. Nevertheless, the radiobiological parameters of the different tumors and normal tissues are often not taken into account. The use of the biologically effective uniform dose (D=) together with the complication-free tumor control probability (P+) were applied to evaluate the two radiation modalities. Two different clinical cases of brain and cranio-spinal axis cancers have been investigated by developing a linac MLC-based step-and-shoot IMRT plan and a Helical Tomotherapy plan. The treatment plans of the MLC-based IMRT were developed on the Philips treatment planning station using the Pinnacle 7.6 software release while the dedicated Tomotherapy treatment planning station was used for the HT plan. With the use of the P+ index and the D = concept as the common prescription point, the different treatment plans were compared based on radiobiological measures. The tissue response probabilities were plotted against D= for a range of prescription doses. The applied plan evaluation method shows that in the brain cancer, the HT treatment gives slightly better results than the MLC-based IMRT in terms of optimum expected clinical outcome (P+ of 66.1{\%} and 63.5{\%} for a D = to the PTV of 63.0 Gy and 62.0 Gy, respectively). In the cranio-spinal axis cancer, the HT plan is significantly better compared to the MLC-based IMRT plan over the clinically useful dose prescription range (P + of 84.1{\%} and 28.3{\%} for a D= to the PTV of 50.6 Gy and 44.0 Gy, respectively). If a higher than 5{\%} risk for complications could be allowed, the complication-free tumor control could be increased by almost 30{\%} compared to the initial dose prescription. In comparison to MLC based-IMRT, HT can better encompass the often large PTV while minimizing the volume of the OARs receiving high dose. A radiobiological treatment plan evaluation can provide a closer association of the delivered treatment with the clinical outcome by taking into account the dose-response relations of the irradiated tumors and normal tissues. The use of P - D= diagrams can complement the traditional tools of evaluation such as DVHs, in order to compare and effectively evaluate different treatment plans.",
keywords = "Biologically effective uniform dose, Complication-free tumor control, Dose-response relations, Helical Tomotherapy, IMRT, Radiobiological evaluation",
author = "Panayiotis Mavroidis and Ferreira, {Brigida Costa} and Chengyu Shi and Delichas, {Miltiadis G.} and Lind, {Bengt K.} and Nikos Papanikolaou",
year = "2009",
month = "2",
language = "English (US)",
volume = "8",
pages = "3--14",
journal = "Technology in Cancer Research and Treatment",
issn = "1533-0346",
publisher = "Adenine Press",
number = "1",

}

TY - JOUR

T1 - Comparison of the helical tomotherapy and MLC-based IMRT radiation modalities in treating brain and cranio-spinal tumors

AU - Mavroidis, Panayiotis

AU - Ferreira, Brigida Costa

AU - Shi, Chengyu

AU - Delichas, Miltiadis G.

AU - Lind, Bengt K.

AU - Papanikolaou, Nikos

PY - 2009/2

Y1 - 2009/2

N2 - The investigation of the clinical efficacy and effectiveness of Intensity Modulated Radiotherapy (IMRT) using Multileaf Collimators (MLC) and Helical Tomotherapy (HT) has been an issue of increasing interest over the past few years. In order to assess the suitability of a treatment plan, dosimetric criteria such as dose-volume histograms (DVH), maximum, minimum, mean, and standard deviation of the dose distribution are typically used. Nevertheless, the radiobiological parameters of the different tumors and normal tissues are often not taken into account. The use of the biologically effective uniform dose (D=) together with the complication-free tumor control probability (P+) were applied to evaluate the two radiation modalities. Two different clinical cases of brain and cranio-spinal axis cancers have been investigated by developing a linac MLC-based step-and-shoot IMRT plan and a Helical Tomotherapy plan. The treatment plans of the MLC-based IMRT were developed on the Philips treatment planning station using the Pinnacle 7.6 software release while the dedicated Tomotherapy treatment planning station was used for the HT plan. With the use of the P+ index and the D = concept as the common prescription point, the different treatment plans were compared based on radiobiological measures. The tissue response probabilities were plotted against D= for a range of prescription doses. The applied plan evaluation method shows that in the brain cancer, the HT treatment gives slightly better results than the MLC-based IMRT in terms of optimum expected clinical outcome (P+ of 66.1% and 63.5% for a D = to the PTV of 63.0 Gy and 62.0 Gy, respectively). In the cranio-spinal axis cancer, the HT plan is significantly better compared to the MLC-based IMRT plan over the clinically useful dose prescription range (P + of 84.1% and 28.3% for a D= to the PTV of 50.6 Gy and 44.0 Gy, respectively). If a higher than 5% risk for complications could be allowed, the complication-free tumor control could be increased by almost 30% compared to the initial dose prescription. In comparison to MLC based-IMRT, HT can better encompass the often large PTV while minimizing the volume of the OARs receiving high dose. A radiobiological treatment plan evaluation can provide a closer association of the delivered treatment with the clinical outcome by taking into account the dose-response relations of the irradiated tumors and normal tissues. The use of P - D= diagrams can complement the traditional tools of evaluation such as DVHs, in order to compare and effectively evaluate different treatment plans.

AB - The investigation of the clinical efficacy and effectiveness of Intensity Modulated Radiotherapy (IMRT) using Multileaf Collimators (MLC) and Helical Tomotherapy (HT) has been an issue of increasing interest over the past few years. In order to assess the suitability of a treatment plan, dosimetric criteria such as dose-volume histograms (DVH), maximum, minimum, mean, and standard deviation of the dose distribution are typically used. Nevertheless, the radiobiological parameters of the different tumors and normal tissues are often not taken into account. The use of the biologically effective uniform dose (D=) together with the complication-free tumor control probability (P+) were applied to evaluate the two radiation modalities. Two different clinical cases of brain and cranio-spinal axis cancers have been investigated by developing a linac MLC-based step-and-shoot IMRT plan and a Helical Tomotherapy plan. The treatment plans of the MLC-based IMRT were developed on the Philips treatment planning station using the Pinnacle 7.6 software release while the dedicated Tomotherapy treatment planning station was used for the HT plan. With the use of the P+ index and the D = concept as the common prescription point, the different treatment plans were compared based on radiobiological measures. The tissue response probabilities were plotted against D= for a range of prescription doses. The applied plan evaluation method shows that in the brain cancer, the HT treatment gives slightly better results than the MLC-based IMRT in terms of optimum expected clinical outcome (P+ of 66.1% and 63.5% for a D = to the PTV of 63.0 Gy and 62.0 Gy, respectively). In the cranio-spinal axis cancer, the HT plan is significantly better compared to the MLC-based IMRT plan over the clinically useful dose prescription range (P + of 84.1% and 28.3% for a D= to the PTV of 50.6 Gy and 44.0 Gy, respectively). If a higher than 5% risk for complications could be allowed, the complication-free tumor control could be increased by almost 30% compared to the initial dose prescription. In comparison to MLC based-IMRT, HT can better encompass the often large PTV while minimizing the volume of the OARs receiving high dose. A radiobiological treatment plan evaluation can provide a closer association of the delivered treatment with the clinical outcome by taking into account the dose-response relations of the irradiated tumors and normal tissues. The use of P - D= diagrams can complement the traditional tools of evaluation such as DVHs, in order to compare and effectively evaluate different treatment plans.

KW - Biologically effective uniform dose

KW - Complication-free tumor control

KW - Dose-response relations

KW - Helical Tomotherapy

KW - IMRT

KW - Radiobiological evaluation

UR - http://www.scopus.com/inward/record.url?scp=60949109725&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=60949109725&partnerID=8YFLogxK

M3 - Article

C2 - 19166237

AN - SCOPUS:60949109725

VL - 8

SP - 3

EP - 14

JO - Technology in Cancer Research and Treatment

JF - Technology in Cancer Research and Treatment

SN - 1533-0346

IS - 1

ER -