Comparison of image layer location among panoramic machines of the same manufacturer

Thomas F. Razmus, Birgit Junfin Glass, William D. McDavid

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

The central plane of the image layer was located in several panoramic machines of the same manufacturer by making radiographs of a spherical test object placed in different positions. A round image indicated that the test object was at the central plane. The inconsistency of the location of the central plane in machines tested indicates the need for establishing a quality assurance test to verify image-layer location before patient exposure.

Original languageEnglish (US)
Pages (from-to)102-108
Number of pages7
JournalOral Surgery, Oral Medicine, Oral Pathology
Volume67
Issue number1
DOIs
StatePublished - 1989

ASJC Scopus subject areas

  • Pathology and Forensic Medicine
  • Dentistry(all)

Cite this

Comparison of image layer location among panoramic machines of the same manufacturer. / Razmus, Thomas F.; Glass, Birgit Junfin; McDavid, William D.

In: Oral Surgery, Oral Medicine, Oral Pathology, Vol. 67, No. 1, 1989, p. 102-108.

Research output: Contribution to journalArticle

Razmus, Thomas F. ; Glass, Birgit Junfin ; McDavid, William D. / Comparison of image layer location among panoramic machines of the same manufacturer. In: Oral Surgery, Oral Medicine, Oral Pathology. 1989 ; Vol. 67, No. 1. pp. 102-108.
@article{dc503d370b7c454f90381bf9992087d0,
title = "Comparison of image layer location among panoramic machines of the same manufacturer",
abstract = "The central plane of the image layer was located in several panoramic machines of the same manufacturer by making radiographs of a spherical test object placed in different positions. A round image indicated that the test object was at the central plane. The inconsistency of the location of the central plane in machines tested indicates the need for establishing a quality assurance test to verify image-layer location before patient exposure.",
author = "Razmus, {Thomas F.} and Glass, {Birgit Junfin} and McDavid, {William D.}",
year = "1989",
doi = "10.1016/0030-4220(89)90311-3",
language = "English (US)",
volume = "67",
pages = "102--108",
journal = "Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology",
issn = "2212-4403",
publisher = "Elsevier USA",
number = "1",

}

TY - JOUR

T1 - Comparison of image layer location among panoramic machines of the same manufacturer

AU - Razmus, Thomas F.

AU - Glass, Birgit Junfin

AU - McDavid, William D.

PY - 1989

Y1 - 1989

N2 - The central plane of the image layer was located in several panoramic machines of the same manufacturer by making radiographs of a spherical test object placed in different positions. A round image indicated that the test object was at the central plane. The inconsistency of the location of the central plane in machines tested indicates the need for establishing a quality assurance test to verify image-layer location before patient exposure.

AB - The central plane of the image layer was located in several panoramic machines of the same manufacturer by making radiographs of a spherical test object placed in different positions. A round image indicated that the test object was at the central plane. The inconsistency of the location of the central plane in machines tested indicates the need for establishing a quality assurance test to verify image-layer location before patient exposure.

UR - http://www.scopus.com/inward/record.url?scp=0024528387&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0024528387&partnerID=8YFLogxK

U2 - 10.1016/0030-4220(89)90311-3

DO - 10.1016/0030-4220(89)90311-3

M3 - Article

C2 - 2911437

AN - SCOPUS:0024528387

VL - 67

SP - 102

EP - 108

JO - Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology

JF - Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology

SN - 2212-4403

IS - 1

ER -