Comparison of continuous transcutaneous and continuous intravascular Po2 measurement in fetal sheep

C. A.M. Jansen, F. G. Bass, K. C. Lowe, P. W. Nathanielsz

Research output: Contribution to journalArticle

8 Scopus citations

Abstract

Continuously monitored carotid arterial Po2 (Pao2) and transcutaneous Po2 (tcPo2) from the scalp of fetuses from ewes not in labor have been compared under epidural anesthesia at different levels of fetal oxygenation. A reasonably good correlation existed between the two readings at normal and high Po2 values. In sharp contrast, at low Pa(o2) levels there was a considerable bias between the two. Since this bias was variable, the failure of tcPo2 to reflect Pa(o2) accurately at low Pa(o2) renders the use of the tcPo2 electrode questionable in the very situation for which it was advocated, namely, the prediction of fetal compromise.

Original languageEnglish (US)
Pages (from-to)670-676
Number of pages7
JournalAmerican Journal of Obstetrics and Gynecology
Volume138
Issue number6
DOIs
StatePublished - Jan 1 1980
Externally publishedYes

ASJC Scopus subject areas

  • Obstetrics and Gynecology

Fingerprint Dive into the research topics of 'Comparison of continuous transcutaneous and continuous intravascular Po<sub>2</sub> measurement in fetal sheep'. Together they form a unique fingerprint.