When significant thoracic vascular anomalies occur in children, they may present surgical difficulties making operative management undesirable. The recent development of a new, accurate coil-delivery system has enabled us to embolize 17 vessels in five children by passing Gianturco steel coils with thrombogenic Dacron strands through No. 5 or 6F end-hole catheters suitable for infants. Coil of 0.038 inch packed diameter were fed through the catheter lumen by a flexible guidewire emerging as 3, 5, or 8 mm diameter loose coils. Fifteen of 17 vessels were successfully occluded. No complications or errors in placement of coils occurred. Four of five children clearly benefitted from the procedure. One died in spite of partial occlusion. Coil embolization can be performed accurately and safety even in small infants with a high rate of successful occlusion and may prove to be a valuable adjunct to operative management.

Original languageEnglish (US)
Pages (from-to)285-289
Number of pages5
JournalCirculation
Volume70
Issue number2
StatePublished - 1984
Externally publishedYes

Fingerprint

Blood Vessels
Thorax
Catheters
Polyethylene Terephthalates
Steel
Safety

ASJC Scopus subject areas

  • Physiology
  • Cardiology and Cardiovascular Medicine

Cite this

Fuhrman, B. P., Bass, J. L., & Castaneda-Zuniga, W. (1984). Coil embolization of congenital thoracic vascular anomalies in infants and children. Circulation, 70(2), 285-289.

Coil embolization of congenital thoracic vascular anomalies in infants and children. / Fuhrman, B. P.; Bass, J. L.; Castaneda-Zuniga, W.

In: Circulation, Vol. 70, No. 2, 1984, p. 285-289.

Research output: Contribution to journalArticle

Fuhrman, BP, Bass, JL & Castaneda-Zuniga, W 1984, 'Coil embolization of congenital thoracic vascular anomalies in infants and children', Circulation, vol. 70, no. 2, pp. 285-289.
Fuhrman BP, Bass JL, Castaneda-Zuniga W. Coil embolization of congenital thoracic vascular anomalies in infants and children. Circulation. 1984;70(2):285-289.
Fuhrman, B. P. ; Bass, J. L. ; Castaneda-Zuniga, W. / Coil embolization of congenital thoracic vascular anomalies in infants and children. In: Circulation. 1984 ; Vol. 70, No. 2. pp. 285-289.
@article{1b6b98eda867496cb31aef3811200844,
title = "Coil embolization of congenital thoracic vascular anomalies in infants and children",
abstract = "When significant thoracic vascular anomalies occur in children, they may present surgical difficulties making operative management undesirable. The recent development of a new, accurate coil-delivery system has enabled us to embolize 17 vessels in five children by passing Gianturco steel coils with thrombogenic Dacron strands through No. 5 or 6F end-hole catheters suitable for infants. Coil of 0.038 inch packed diameter were fed through the catheter lumen by a flexible guidewire emerging as 3, 5, or 8 mm diameter loose coils. Fifteen of 17 vessels were successfully occluded. No complications or errors in placement of coils occurred. Four of five children clearly benefitted from the procedure. One died in spite of partial occlusion. Coil embolization can be performed accurately and safety even in small infants with a high rate of successful occlusion and may prove to be a valuable adjunct to operative management.",
author = "Fuhrman, {B. P.} and Bass, {J. L.} and W. Castaneda-Zuniga",
year = "1984",
language = "English (US)",
volume = "70",
pages = "285--289",
journal = "Circulation",
issn = "0009-7322",
publisher = "Lippincott Williams and Wilkins",
number = "2",

}

TY - JOUR

T1 - Coil embolization of congenital thoracic vascular anomalies in infants and children

AU - Fuhrman, B. P.

AU - Bass, J. L.

AU - Castaneda-Zuniga, W.

PY - 1984

Y1 - 1984

N2 - When significant thoracic vascular anomalies occur in children, they may present surgical difficulties making operative management undesirable. The recent development of a new, accurate coil-delivery system has enabled us to embolize 17 vessels in five children by passing Gianturco steel coils with thrombogenic Dacron strands through No. 5 or 6F end-hole catheters suitable for infants. Coil of 0.038 inch packed diameter were fed through the catheter lumen by a flexible guidewire emerging as 3, 5, or 8 mm diameter loose coils. Fifteen of 17 vessels were successfully occluded. No complications or errors in placement of coils occurred. Four of five children clearly benefitted from the procedure. One died in spite of partial occlusion. Coil embolization can be performed accurately and safety even in small infants with a high rate of successful occlusion and may prove to be a valuable adjunct to operative management.

AB - When significant thoracic vascular anomalies occur in children, they may present surgical difficulties making operative management undesirable. The recent development of a new, accurate coil-delivery system has enabled us to embolize 17 vessels in five children by passing Gianturco steel coils with thrombogenic Dacron strands through No. 5 or 6F end-hole catheters suitable for infants. Coil of 0.038 inch packed diameter were fed through the catheter lumen by a flexible guidewire emerging as 3, 5, or 8 mm diameter loose coils. Fifteen of 17 vessels were successfully occluded. No complications or errors in placement of coils occurred. Four of five children clearly benefitted from the procedure. One died in spite of partial occlusion. Coil embolization can be performed accurately and safety even in small infants with a high rate of successful occlusion and may prove to be a valuable adjunct to operative management.

UR - http://www.scopus.com/inward/record.url?scp=0021131426&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0021131426&partnerID=8YFLogxK

M3 - Article

C2 - 6733882

AN - SCOPUS:0021131426

VL - 70

SP - 285

EP - 289

JO - Circulation

JF - Circulation

SN - 0009-7322

IS - 2

ER -