T1 - Biomechanical comparison of the sandblasted and acid-etched and the machined and acid-etched titanium surface for dental implants
AU - Li, Dehua
AU - Ferguson, Stephen J.
AU - Beutler, Thomas
AU - Cochran, David L.
AU - Sittig, Caroline
AU - Hirt, Hans Peter
AU - Buser, Daniel
PY - 2002
Y1 - 2002
N2 - To make a direct biomechanical comparison between the sandblasted and acid-etched surface (SLA) and the machined and acid-etched surface (MA), a well-established animal model for implant removal torque testing was employed, using a split-mouth experimental design. All implants had an identical cylindrical solid-screw shape with the standard ITI thread configuration, without any macroscopic retentive structures. After 4, 8, and 12 weeks of bone healing, removal torque testing was performed to evaluate the interfacial shear strength of each surface type. Results showed that the SLA surface was more powerful in enhancing the interfacial shear strength of implants in comparison with the MA surface. Removal torque values of the SLA-surfaced implants were about 30% higher than those of the MA-surfaced implants (p = 0.002) except at 4 weeks, when the difference was at the threshold of statistical significance (p = 0.0519). The mean removal torque values for the SLA implants were 1.5074 Nm at 4 weeks, 1.8022 Nm at 8 weeks, and 1.7130 Nm at 12 weeks; and correspondingly, 1.1924 Nm, 1.3092 Nm, and 1.3226 Nm for the MA implants. It can be concluded that the SLA surface achieves a better bone anchorage than the MA surface, and that sandblasting before acid etching has a beneficial effect on the interfacial shear strength. As regards the bone-implant interfacial stiffness calculated from the torque-rotation curve, the SLA implants showed an overall more than 5% higher stiffness compared with the MA implants, although the difference did not reach the statistical significance level.
AB - To make a direct biomechanical comparison between the sandblasted and acid-etched surface (SLA) and the machined and acid-etched surface (MA), a well-established animal model for implant removal torque testing was employed, using a split-mouth experimental design. All implants had an identical cylindrical solid-screw shape with the standard ITI thread configuration, without any macroscopic retentive structures. After 4, 8, and 12 weeks of bone healing, removal torque testing was performed to evaluate the interfacial shear strength of each surface type. Results showed that the SLA surface was more powerful in enhancing the interfacial shear strength of implants in comparison with the MA surface. Removal torque values of the SLA-surfaced implants were about 30% higher than those of the MA-surfaced implants (p = 0.002) except at 4 weeks, when the difference was at the threshold of statistical significance (p = 0.0519). The mean removal torque values for the SLA implants were 1.5074 Nm at 4 weeks, 1.8022 Nm at 8 weeks, and 1.7130 Nm at 12 weeks; and correspondingly, 1.1924 Nm, 1.3092 Nm, and 1.3226 Nm for the MA implants. It can be concluded that the SLA surface achieves a better bone anchorage than the MA surface, and that sandblasting before acid etching has a beneficial effect on the interfacial shear strength. As regards the bone-implant interfacial stiffness calculated from the torque-rotation curve, the SLA implants showed an overall more than 5% higher stiffness compared with the MA implants, although the difference did not reach the statistical significance level.
KW - Acid-etched surface
KW - Interfacial shear strength
KW - Interfacial stiffness
KW - Removal torque values
KW - Sandblasted and acid-etched surface
KW - Titanium implants
UR - http://www.scopus.com/inward/record.url?scp=0036185427&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0036185427&partnerID=8YFLogxK
U2 - 10.1002/jbm.10063
DO - 10.1002/jbm.10063
M3 - Article
C2 - 11857440
AN - SCOPUS:0036185427
SN - 0021-9304
VL - 60
SP - 325
EP - 332
JO - Journal of Biomedical Materials Research
JF - Journal of Biomedical Materials Research
IS - 2
ER -